Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet Spine, LLC: MaxAn 4.0mm and 4.5mm fixed screws, used in the MaxAn Anterior Cervical Plate System. The MaxAn Anterior Cervical Plate System is a decompression-based technique for cervical spine stabilization.

Recall number
Z-0203-2016
Recalling firm
Biomet Spine, LLC
Product
MaxAn 4.0mm and 4.5mm fixed screws, used in the MaxAn Anterior Cervical Plate System. The MaxAn Anterior Cervical Plate System is a decompression-based technique for cervical spine stabilization.
Status
Terminated
Initiated
2015-09-29
Posted
2015-11-03
Terminated
2016-04-29
Reason
The packaging of 4.5 mm diameter MaxAn screws identifies/labels the screws as 4.0 mm diameter screws.
Root cause
Labeling mix-ups
Action
Zimmer Biomet Spine sent an Urgent Medical Device Recall Notice dated September 29, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. If the device has already been implanted, written confirmation of this will be required and the necessary parties will be contacted. They will store returned product in quarantine in an isolated location until they can undergo inspection to verify if they are conforming or not and then will be destroyed or retained. Necessary parts will be destroyed so that it is visually obvious they are no longer functional. For further questions call (303) 443-7500 ext.244.
Quantity
17 units
Distribution
US Distribution to the states of : NY, MO, and FL
Lot, serial or model codes
P/N: 14-521550; Lot Numbers: 463870;  P/N: 14-521548; Lot Numbers: J86056;  P/N: 14-521518; Lot Numbers: J26394; Model Numbers: 14-521518 14-521520 14-521548 14-521550
510(k) numbers
["K080646","K133518"]
PMA numbers
not on file
Product code
KWQ
Firm FEI
3012447612
Firm city
Broomfield
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register