Ethicon Endo-Surgery Inc: ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 440mm shaft, Product Code PLEE60A

Recall number
Z-0202-2020
Recalling firm
Ethicon Endo-Surgery Inc
Product
ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 440mm shaft, Product Code PLEE60A
Status
Terminated
Initiated
2019-10-03
Posted
2019-10-29
Terminated
2021-06-03
Reason
The staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple line is compromised, there is a potential risk of prolonged surgery, postoperative anastomotic leak, hemorrhage, hemorrhagic shock, additional surgical intervention, or death.
Root cause
Nonconforming Material/Component
Action
Urgent Medical Device Recall (Removal) letters were sent to customers beginning 10/3/19. The letters instruct customers to do the following: ACTION REQUIRED: 1. Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2. Remove the product subject to this voluntary recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 4. Complete the Business Reply Form (BRF) (Attachment 2) confirming receipt of this notice and fax or email it to Stericycle at 1-866-300-3432 or ethicon7148@stericycle.com within three (3) business days. Please return the BRF even if you do not have product subject to this recall. 5. Keep this notice visibly posted for awareness until all product subject to this recall has been returned to Stericycle. While processing your returns, please maintain a copy of this notice with the product subject to this recall and keep a copy for your records. 6. Customers are required to return unused impacted ECHELON FLEX ENDOPATH 60mm Staplers subject to this recall that are in their inventory immediately. To receive replacement product, customers must return product subject to this recall by December 31, 2019. Any nonaffected product and any product returned after the date specified will not be replaced. 7. To return product subject to this recall, photocopy the completed BRF, place it in the box with the product, and affix the pre-paid authorized shipping label included with the recall notification letter. Ethicon will pay for the shipping charges only if the authorized label is used. Extra shipping labels may be obtained by calling Stericycle at 1-866-918-8756. Your account n
Quantity
5040 units
Distribution
Worldwide distribution. US Nationwide. Belgium, Brazil, Canada, Chile, Colombia, Japan, Kuwait, Saudi Arabia, and Singapore.
Lot, serial or model codes
Lot: T93Z1G, T93Y4M, T94045, T94117, T93X17, T94087, T93Z2X, T9405L, T94253, T93X95, T93Z75, T93Z2W, T9413Z
510(k) numbers
["K140560"]
PMA numbers
not on file
Product code
GDW
Firm FEI
1527736
Firm city
Blue Ash
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register