FDA Medical Device Recalls (openFDA)
Integra LifeSciences Corp.: Integra Licox Pt02 Monitor, Rx Only, Manufacturer: Integra LifeSciences (Ireland) Limited, IDA Business and Technology Park, Sragh, Tullamore, County Offaly, Ireland, Distributed by: Integra LifeSciences Corporation, 311 Enterprise Drive, Plainsboro, NJ 08536
- Recall number
- Z-0201-2015
- Recalling firm
- Integra LifeSciences Corp.
- Product
- Integra Licox Pt02 Monitor, Rx Only, Manufacturer: Integra LifeSciences (Ireland) Limited, IDA Business and Technology Park, Sragh, Tullamore, County Offaly, Ireland, Distributed by: Integra LifeSciences Corporation, 311 Enterprise Drive, Plainsboro, NJ 08536
- Status
- Terminated
- Initiated
- 2014-08-18
- Posted
- 2014-11-10
- Terminated
- 2016-05-11
- Reason
- Integra LifeSciences Corporation received 2 complaints that the USB port on the Licox Pt02 monitor does not consistently provide the user the ability to extract the Pt02 trend data according to the User's Manual for the device.
- Root cause
- Device Design
- Action
- Integra LifeSciences sent an "Urgent Voluntary Medical Device Correction" letter/Return Response Form dated August 18, 2014, to their affected customers via traceable courier service. The letter identified the product the problem and the action needed to be taken by the customer. Complete the attached form. If you do not have an affected monitor, check the box: I do not have any monitors with an affected Serial Number listed below. If you do have affected monitors, check the box: I do have monitors with an affected Serial Number listed below. Circle all the Serial Numbers for the affected monitors you have and record the total quantity. Complete the other information on the form and return the form by email or fax as indicated. Keep a copy of the form for your records. When your form is received, Integra will schedule this service with your facility and provide a temporary loaner monitor to minimize any disruption this action may cause. Should you have any questions, please contact Neuro Customer Service at 1-855-532-1723. We apologize for any inconvenience this may cause and thank you for your cooperation in this effort.
- Quantity
- 43 (LCX02) and 17 (LCX02R)
- Distribution
- US Distribution including the states of : CA, DC, FL, IL, IN, MI, MN, NC, NY, PA and TX
- Lot, serial or model codes
- Catalogue #LCX02 - Serial #'s 2130901032, 2130901050, 2131001062, 2131001063, 2131001086, 2131001087, 2131001102, 2131001110, 2131001111, 2131001115, 2131101124, 2131101125, 2131101127, 2131101129, 2131101134, 2131101138, 2131101151, 2131101154, 2140101181, 2140301222, 2130901017, 2130901019, 2130901022, 2130901025, 2130901034, 2130901035, 2130901051, 2130901055, 2130901056, 2130901058, 2130901059, 2131001072, 2131001076, 2131001077, 2131001081, 2131001088, 2131001093, 2131001108, 2131101126, 2131101129, 2131101122, 2130901028 and 2130901039; Catalogue #LCX02R - Serial #'s 2130901038, 2130901046, 2130901054, 2131201163, 2131201164, 2131201165, 2131201166, 2131201167, 2140101170, 2140101172, 2140101174, 2140101178, 2140101179, 2140201186, 2140201189, 2140201197 and 2140401227.
- 510(k) numbers
- ["K131184"]
- PMA numbers
- not on file
- Product code
- GWM
- Firm FEI
- 3003418325
- Firm city
- Plainsboro
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28