FDA Medical Device Recalls (openFDA)
Toshiba American Medical Systems Inc: Titan 3T MRI System, MRT-3010A/5. Product Usage: Usage: MRI system.
- Recall number
- Z-0201-2014
- Recalling firm
- Toshiba American Medical Systems Inc
- Product
- Titan 3T MRI System, MRT-3010A/5. Product Usage: Usage: MRI system.
- Status
- Terminated
- Initiated
- 2013-10-28
- Posted
- 2013-11-13
- Terminated
- 2014-12-24
- Reason
- Toshiba America Medical Systems, Inc. is recalling due to a potential problem with Specific Absorption Rate (SAR) values when using the Toshiba Titan 3T System. When the system is used in combination with either the QD Head Coil (MJQH-142A) or the Large Knee SPEEDER (MJAJ-182A), SAR values lower than the actual SAR values may be displayed on the operating monitor.
- Root cause
- Nonconforming Material/Component
- Action
- Toshiba America Medical Systems, Inc. sent an "URGENT: MEDICAL DEVICE CORRECTION". letter dated October 28, 2013 via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS for retention. The letter listed Device, Serial Number, reason for recall, problem, corrective actions, request to customers, and contact information.
- Quantity
- 18 units
- Distribution
- USA Nationwide Distribution in the states of: CA, FL, GA, IA, NV, NY, LA, IN, PA, OR, TX, and WI.
- Lot, serial or model codes
- Lot numbers not utilized. Serial Numbers:A5B1192003 A5B1192004, A5B11Y2005, A5B11Z2007, A5C1212008, A5C1212009, A5C1222010, A5C1232011, A5C1252012, A5C1262013, A5C1282014, A5C1292015, A5C12X2016, A5C12Y2017, A5C12Z2018, A5D1312019, A5D1322020, A5D1362021, XIB-A5A1082001, XIB-A5B11Y2006, A5D1372022, A5B1182002.
- 510(k) numbers
- ["K102489"]
- PMA numbers
- not on file
- Product code
- LNH
- Firm FEI
- 2020563
- Firm city
- Tustin
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28