Randox Laboratories Ltd.: RX Imola-IVD Clinical Chemistry Analyzer Catalog Number: RX4900. RX4900R (refurbished)
- Recall number
- Z-0200-2024
- Recalling firm
- Randox Laboratories Ltd.
- Product
- RX Imola-IVD Clinical Chemistry Analyzer Catalog Number: RX4900. RX4900R (refurbished)
- Status
- Open, Classified
- Initiated
- 2023-09-06
- Posted
- 2023-10-31
- Terminated
- not on file
- Reason
- RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
- Root cause
- Under Investigation by firm
- Action
- Randox issued Medical Device Correction letter (REC687) on 9/13/23 to to US and PR customers on 9/12/23. Letter states reason for recall, health risk and action to take: Randox recommendations: " Auto rerun settings are disabled, see Appendix A for guidance for each Rx instrument. " Results with flag(<) reported as the bottom of the technical range. Sample Volume settings Low should not be applied. " Results with flag (>) reported as the top of the technical range. Sample Volume settings high and dilutions should not be applied. A manual dilution should be performed, and results multiplied by the dilution factor. The RX Parameter rerun settings will be updated with a parameter release in due course, the Instructions For Use will also be updated at this time. Action to be taken: " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation.
- Quantity
- 35 units
- Distribution
- US Nationwide distribution including Puerto Rico
- Lot, serial or model codes
- GTIN: 05055273206104 Lot/Batch Number: (1) 7201-0747 7201-0423 7201-0417 7201-0353 7255-0198 7255-0172 7255-0171 7255-0129 7201-1085 7201-1086 7201-1079 7201-1080 7201-0982 7201-0977 7201-0970 7201-0950 7201-0924 7201-0919 7201-0901 7201-0849 7201-0827 7201-0801 7201-0768 7201-0606 7201-0562 7201-0535 7201-0531 7201-0482 (2)¿¿¿¿¿¿¿¿¿¿¿¿¿ 7201-0776 7201-0367 7201-0343¿¿¿¿¿¿¿¿¿¿¿¿¿ ¿¿¿¿ 7201-0348 ¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿
- 510(k) numbers
- ["K052914"]
- PMA numbers
- not on file
- Product code
- CEM
- Firm FEI
- 1000361607
- Firm city
- Crumlin
- Firm state
- North
- Firm country
- Ireland
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28