Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare: Monitor suspensions used with Advantx Legacy, RFX Classical, SFX-90, Model # 46-240838G2, manufactured after March 8, 1985 for Advantx Legacy, RFX and SFX systems. General-purpose radiographic system component used to suspend the system's image monitor.

Recall number
Z-0200-2009
Recalling firm
GE Healthcare
Product
Monitor suspensions used with Advantx Legacy, RFX Classical, SFX-90, Model # 46-240838G2, manufactured after March 8, 1985 for Advantx Legacy, RFX and SFX systems. General-purpose radiographic system component used to suspend the system's image monitor.
Status
Terminated
Initiated
2007-03-09
Posted
2008-10-27
Terminated
2012-08-24
Reason
GE Healthcare has recently become aware of a potential risk of monitor dropping associated with the monitor suspension of the Advantx system that may impact patient safety. It has been reported that suspensions manufactured after March 8, 1995 did not have the required thread locking agent applied to the setscrew, allowing the setscrew to back out over time, which can lead to the monitor dropping
Root cause
Component change control
Action
Consignees were sent a GE Healthcare "Product Safety Notification" letter dated September 13, 2007. The letter was addressed to Hospital Administrators/Managers of Radiology/Radiologists. The letter described the Safety Issue, Affected Product, Safety Instructions, Corrective Action and Contact Information. The letter recommends that customers move the monitor suspension to an open area in the room where viewing is still possible and make sure to minimize the movement of the monitor until a Qualified GE Healthcare Field Engineer has inspected the suspension. The letter also states that a Qualified GE Healthcare Field Engineer will schedule a field visit to inspect the monitor suspension. Customers should contact their sales or service rep is they have any questions.
Quantity
21
Distribution
Worldwide Distribution --- USA including states of AZ, CA, FL, GA, IA, MD, MO, NC, NY, OH, OK, TN, TX,WI, and LA, and country of Slovak Republic.
Lot, serial or model codes
00000486566WK3, 00000486586WK1, 00000486555WK6, 00000486573WK9, 00000486581WK2, 00000478555WK6, 00000486579WK6, 00000486588WK7, 00000486557WK2, 00000486560WK6, 00000486550WK7, 00000486556WK4, 00000486547WK3, 00000056366VP8, 00000486544WK0, 00000486541WK6, 00000486543WK2, 00000471291WK5, 00000486568WK9, and 00000486559WK8
510(k) numbers
["K760466","K801320","K931049"]
PMA numbers
not on file
Product code
IZO
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28