Randox Laboratories Ltd.: RX Daytona Plus (with ISE)-IVD Clinical Chemistry Analyser Catalog Number: RX4040

Recall number
Z-0198-2024
Recalling firm
Randox Laboratories Ltd.
Product
RX Daytona Plus (with ISE)-IVD Clinical Chemistry Analyser Catalog Number: RX4040
Status
Open, Classified
Initiated
2023-09-06
Posted
2023-10-31
Terminated
not on file
Reason
RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
Root cause
Under Investigation by firm
Action
Randox issued Medical Device Correction letter (REC687) on 9/13/23 to to US and PR customers on 9/12/23. Letter states reason for recall, health risk and action to take: Randox recommendations: " Auto rerun settings are disabled, see Appendix A for guidance for each Rx instrument. " Results with flag(<) reported as the bottom of the technical range. Sample Volume settings Low should not be applied. " Results with flag (>) reported as the top of the technical range. Sample Volume settings high and dilutions should not be applied. A manual dilution should be performed, and results multiplied by the dilution factor. The RX Parameter rerun settings will be updated with a parameter release in due course, the Instructions For Use will also be updated at this time. Action to be taken: " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation.
Quantity
5 units
Distribution
US Nationwide distribution including Puerto Rico
Lot, serial or model codes
GTIN: 05055273207774 Lot/ Batch Number (1) Lot/ Batch Number 7241-0551 7241-0545 7241-0531 7241-0317 7241-0219
510(k) numbers
["K024014"]
PMA numbers
not on file
Product code
CGA
Firm FEI
1000361607
Firm city
Crumlin
Firm state
North
Firm country
Ireland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register