Skip to the content

FDA Medical Device Recalls (openFDA)

Becton Dickinson & Co.: BD PhoenixTM AP, Catalog Number 448010 The BD Phoenix" AP instrument is designed for use with the BD Phoenix" system is intended to standardize ID broth tube inoculum, add AST indicator to the AST broth tube and transfer an aliquot of ID broth to AST broth tubes, as required for preparing samples for use on the BD Phoenix" system, which performs identification and susceptibility testing

Recall number
Z-0198-2018
Recalling firm
Becton Dickinson & Co.
Product
BD PhoenixTM AP, Catalog Number 448010 The BD Phoenix" AP instrument is designed for use with the BD Phoenix" system is intended to standardize ID broth tube inoculum, add AST indicator to the AST broth tube and transfer an aliquot of ID broth to AST broth tubes, as required for preparing samples for use on the BD Phoenix" system, which performs identification and susceptibility testing
Status
Terminated
Initiated
2017-06-01
Posted
not on file
Terminated
2018-05-09
Reason
Potential unexpected movement of robot arm
Root cause
Process control
Action
BD Diagnostic Systems telephoned all customers on June 1, 2017 and sent an Urgent Product Recall letter dated May 31, 2017, and Response Forms to their customers via email. BD recently confirmed a manufacturing issue with the motor that drives movement of the robot arm. The issue may result in the inability of this motor to withstand the lateral forces encountered during normal operations. Should this occur, the instrument will indicate an E20 error-Pipettor step loss during operation and stop functioning. While there is no loss of patient sample, the samples would need to be manually processed until the instrument can be repaired. To limit any disruption in the workload, BD Service will contact customers to schedule the instrument replacement. Customers are advised to complete and return the response form via email to BDRC2@aol.com. Customers may continue to use the instrument until replacement is performed. Customers with questions may contact BD (Customer/Technical) Support at 1-800-638-8663. For questions regarding this recall call 410-316-4258.
Quantity
7 units
Distribution
Worldwide Distribution - US including CA, NC & TN Internationally: Canada
Lot, serial or model codes
Serial Numbers: AP0447, AP0448, AP0449, AP0450, AP0451, AP0458 & AP0463.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQX
Firm FEI
1119779
Firm city
Sparks
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register