Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diabetes Care, Inc.: Accu-Chek Connect diabetes management software app, Catalog number 07562462001 / GTIN number 00365702700000, Catalog number 07250452001 / GTIN number 00365702700017 The Accu-Chek Connect Diabetes Management App is indicated as an aid in the treatment of diabetes. The software provides for electronic download of blood glucose meters, manual data entry, storage, display, transfer, and self-managing of blood glucose and other related health indicators which can be shown in report and graphical format. The Accu-Chek Bolus Advisor, as a component of the Accu-Chek Connect Diabetes Management App, is indicated for the management of diabetes by calculating an insulin dose or carbohydrate intake based on user-entered data

Recall number
Z-0197-2018
Recalling firm
Roche Diabetes Care, Inc.
Product
Accu-Chek Connect diabetes management software app, Catalog number 07562462001 / GTIN number 00365702700000, Catalog number 07250452001 / GTIN number 00365702700017 The Accu-Chek Connect Diabetes Management App is indicated as an aid in the treatment of diabetes. The software provides for electronic download of blood glucose meters, manual data entry, storage, display, transfer, and self-managing of blood glucose and other related health indicators which can be shown in report and graphical format. The Accu-Chek Bolus Advisor, as a component of the Accu-Chek Connect Diabetes Management App, is indicated for the management of diabetes by calculating an insulin dose or carbohydrate intake based on user-entered data
Status
Terminated
Initiated
2017-06-08
Posted
not on file
Terminated
2018-08-06
Reason
Certain iOS and Android App versions contain a program error (bug) in the Bolus Advisor feature. After pairing a meter with the app for the first time, a customer may encounter the rare condition in which the countdown timer is not displayed and correction bolus advice is not available for the most recent, valid glucose reading. This same blood glucose value may become available for bolus advice calculation at a later time (countdown timer is displayed).
Root cause
Software design
Action
Roche Diabetes Care sent an Urgent Medical Device Correction letter dated June 12, 2017, to all affected customers. The potential for inaccurate bolus insulin advice can only be encountered if the blood glucose meter has displayed a single E-10 Time/Date error message and the meter had blood glucose records in My Diary and then the meter is pared for the first time with the Accu-Chek Connect App. A customer encountering this issue may find the app does not display the bolus advice option and, consequently, correction bolus advice is not available on the most recent, valid glucose reading. This blood glucose value that could not be used initially may become available for bolus advice calculation at a later time. If the app provides the bolus advice option more than 15 minutes after you tested, it could lead to an incorrect bolus insulin recommendation that should not be used. To completely avoid the possibility of this issue occurring, it is important for Accu-Chek Connect app users to follow the following directions when pairing a meter with the app for the first time: Clear all data from your Accu-Chek blood glucose meter following the steps below, before pairing with the Accu-Chek Connect app for the first time. This clearing of the records in the memory can be accomplished on your PC or Mac by using Accu-Chek Connect Device Link. A micro-USB cable and an Accu-Chek Connect Online account are required. Also ensure that the date and time are set correctly on your blood glucose meter before pairing with the app. Failure to follow these directions might impact first time use of the apps Bolus Advisor feature. Steps to clear data from your Accu-Chek meter in Connect Online: 1. Create or login to your Connect Online account on your PC or Mac. 2. Download Accu-Chek Device Link 3. Right-click on Device Link icon in system tray or dock. 4. Select Settings >Prompt after every successful upload under the Delete Patient Results from Device heading. 5. Connect y
Quantity
1,134 bolus advisor features used
Distribution
Worldwide Distribution - US (Nationwide) and Canada
Lot, serial or model codes
Software Versions 1.0.5, 1.0.15, 1.2.0, 1.2.2, 1.2.3, 2.0.0, 2.0.1, 2.1.0, and 2.1.1 for iOS and Android.
510(k) numbers
["K150910"]
PMA numbers
not on file
Product code
NDC
Firm FEI
3011393376
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register