Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: GE Healthcare Optima XR 200amx with digital upgrade Mobile general purpose radiographic imaging of the human head and body.

Recall number
Z-0197-2013
Recalling firm
GE Healthcare, LLC
Product
GE Healthcare Optima XR 200amx with digital upgrade Mobile general purpose radiographic imaging of the human head and body.
Status
Terminated
Initiated
2012-09-20
Posted
2012-11-16
Terminated
2013-12-05
Reason
It was discovered GE Healthcare has become aware of a firmware issue on the mobile radiographic product, Brivo XR285amx, Optima XR200amx, Optima XR22Oamx and Optima XR200amx with Digital Upgrade, related to Technique Accuracy as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-2-54.
Root cause
Radiation Control for Health and Safety Act
Action
GE Healthcare will remedy the defect or bring the product into compliance with each applicable Federal and IEC standard. A Service Representative will update the firmware on the system to address the issue. Field Modification Instructions (FMI) 10870, 10872, and 10873 are describing the rework plan the GEHC Engineers will make to the imaging systems compliant with the applicable performance standard. Further questions please call (262) 513-4122.
Quantity
364 total units installed in US
Distribution
Nationwide Distribution
Lot, serial or model codes
part number: 5421698
510(k) numbers
not on file
PMA numbers
not on file
Product code
OXO
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register