Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Instruments Div. of Stryker Corporation: Right/Left IBO Blade Set, for use with Stryker TPS Reciprocating Saw, REF 5100-37-903, Stryker Instruments, Kalamazoo, MI. Product Usage: The intended use of these devices is to cut bone and hard tissue in maxillofacial procedures and have been specifically designed for osteotomies of the mandibular ramus.

Recall number
Z-0197-2012
Recalling firm
Stryker Instruments Div. of Stryker Corporation
Product
Right/Left IBO Blade Set, for use with Stryker TPS Reciprocating Saw, REF 5100-37-903, Stryker Instruments, Kalamazoo, MI. Product Usage: The intended use of these devices is to cut bone and hard tissue in maxillofacial procedures and have been specifically designed for osteotomies of the mandibular ramus.
Status
Terminated
Initiated
2011-10-18
Posted
2011-11-16
Terminated
2013-07-12
Reason
There is a potential for these blades to break where the blade is welded on to the arbor. This hazard can result in the following harms: tissue damage due to the broken blade being aspirated or swallowed, additional surgery to remove a broken blade, infection due to an unretrieved device component, complications associated with increased time under anesthesia to retrieve a broken blade, osteomyeli
Root cause
Device Design
Action
The firm, Stryker, sent an "URGENT MEDICAL DEVICE RECALL NOTIFICATION" letter dated October 18, 2011 to Risk Managers, Office Managers, and Oral Surgeons. The letter described the products under recall, the problem and a series of actions to be taken. Customers were asked to immediately locate recalled products and quarantine. Products were to be shipped back to Stryker. The letter included a Business Reply Form to be faxed back to Stryker Instruments Regulatory Department, 866-521-2762. The letter instructs to forward this notification to all affected locations if the recalled products were further distributed. Customers with questions can call Stryker at 1-800-253-3210.
Quantity
1,086 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Brazil, Canada, China, Germany, Greece, Hong Kong, India, Jordan, Miramar, Saudi Arabia, Singapore, Spain, Switzerland, United Kingdom.
Lot, serial or model codes
06216017, 06230017, 06247017, 06313017, 06354017, 07067017, 07110017, 07183017, 07190017, 07225017, 07246017, 07323017, 08141017, 08269017, 08336017, 09021017, 09050017, 09083017, 09135017, 09317017, 10085017, 10091017, 10117017, 10148017, 10194017, 10217017, 10327017, 11010017, 11100017, 11119017, 11159017, 11167017,
510(k) numbers
["K040369"]
PMA numbers
not on file
Product code
DZJ
Firm FEI
1811755
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register