FDA Medical Device Recalls (openFDA)
Beckman Coulter Inc: Cytomics FC 500 Flow Cytometry System with CXP Software Versions 2.0 & 2.1
- Recall number
- Z-0197-2007
- Recalling firm
- Beckman Coulter Inc
- Product
- Cytomics FC 500 Flow Cytometry System with CXP Software Versions 2.0 & 2.1
- Status
- Terminated
- Initiated
- 2006-07-26
- Posted
- 2006-11-21
- Terminated
- 2012-02-29
- Reason
- Mis-identification-If a panel or protocol is added to an existing worklist but the tube location is not specified, the CXP Acquisition software will run the last specified tube through the remaining protocols and will generate results and printouts of these runs leading to a Mis-Identification condition.
- Root cause
- Other
- Action
- Customer letter -- A Product Corrective Action (PCA) letter was sent the week of July 24, 2006 to all accounts that have FC500 with CXP Software 2.0 and 2.1 to inform them that a potential sample misidentification may occur when running a Work list with missing tube Location(s) in Automatic MCL mode. The letter outlines actions to be taken to avoid mis-identification of patient results.
- Quantity
- 416 (US) + 60 (Canada)
- Distribution
- Nationwide and Canada
- Lot, serial or model codes
- Part Numbers: 722645 Software Kit; CXP version 2.1 New User; 723113 Software Kit; CXP version 2.1 Upgrade; 629637 Software Kit; CXP version 2.0 New User; 629636 Software Kit; CXP version 2.0 Upgrade; 626553 TN; FC500 WITH UPS; 6605627 TN; 5 CLR, FC500 (IOOV); 6605628 TN; 5 CLR, FC500 (120V); 6605629 TN; 5 CLR, FC500 (220V); 6605630 TN;5 CLR, FC500 (240V)
- 510(k) numbers
- ["K030828"]
- PMA numbers
- not on file
- Product code
- GKZ
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as