Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: Baxter 15L Cycler Drainage Bag Sterile, nonpyrogenic fluid path

Recall number
Z-0194-2016
Recalling firm
Baxter Healthcare Corp.
Product
Baxter 15L Cycler Drainage Bag Sterile, nonpyrogenic fluid path
Status
Terminated
Initiated
2015-10-06
Posted
not on file
Terminated
2016-06-27
Reason
Leakage from the inlet port of the 15L Cycler Drainage Bags.
Root cause
Under Investigation by firm
Action
Baxter sent an Urgent Product Recall letter dated October 8, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The recalling firm will ask customers to locate and separate all affected products, contact the recalling firm to arrange for a return, and complete the customer reply form. Further questions, please call (800) 422-9837.
Quantity
19,440 units
Distribution
US Distribution and Internationally to Canada.
Lot, serial or model codes
H15C11095
510(k) numbers
["K791899"]
PMA numbers
not on file
Product code
KDJ
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register