Skip to the content

FDA Medical Device Recalls (openFDA)

Clinical Diagnostic Solutions: Boule Con-Diff Hematology Control 9 vials. For in vitro diagnostic use as a control to monitor the performance of multi-parameter hematology instruments.

Recall number
Z-0194-2013
Recalling firm
Clinical Diagnostic Solutions
Product
Boule Con-Diff Hematology Control 9 vials. For in vitro diagnostic use as a control to monitor the performance of multi-parameter hematology instruments.
Status
Terminated
Initiated
2010-08-04
Posted
2012-11-05
Terminated
2014-08-14
Reason
Clinical Diagnostic Solutions, Inc. in Plantation, FL recalled the Boule Con Diff US Tri Pack (Product part 501-605, Lot #1005-549 and Product part 501-607, Lot #1005-549) which is used for in vitro diagnostic use as a control to monitor the performance of multi-parameter hematology instruments. The recall was imitated due to receipt of customers stating the MCV values and RDW% recovered for Boule
Root cause
Process control
Action
Clinical Diagnostic Solutions, Inc. sent a "PRODUCT NOTIFICATION:BOULE CON DIFF TRI-LEVEL LOT 1005-549" letter dated August 4, 2010 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. For questions regarding this notice, contact CDS Customer Service at 1-800-453-3328.
Quantity
690 packs
Distribution
Nationwide Distribution.
Lot, serial or model codes
Product part 501-607, Lot #1005-549
510(k) numbers
["K024173"]
PMA numbers
not on file
Product code
JPK
Firm FEI
3002684533
Firm city
Plantation
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register