FDA Medical Device Recalls (openFDA)
Baxter Healthcare Corp.: Baxter Clearlink Intravenous (IV) Delivery System; the system consists of solution sets, extension sets and blood sets that have a luer activated valve for IV access, which allows the administration of medications/solutions with a needleless luer syringe; Baxter Healthcare Corporation, Deerfield, IL 60015 USA
- Recall number
- Z-0193-05
- Recalling firm
- Baxter Healthcare Corp.
- Product
- Baxter Clearlink Intravenous (IV) Delivery System; the system consists of solution sets, extension sets and blood sets that have a luer activated valve for IV access, which allows the administration of medications/solutions with a needleless luer syringe; Baxter Healthcare Corporation, Deerfield, IL 60015 USA
- Status
- Terminated
- Initiated
- 2004-04-28
- Posted
- 2004-11-16
- Terminated
- 2007-02-08
- Reason
- Inadequate directions for use may have contributed to an increase in infection rates at a small number of customers beginning use of the Clearlink needleless device.
- Root cause
- Other
- Action
- An Urgent Device Correction letter and reply form dated 4/28/04 were sent to all U.S. Clearlink customers, informing them of the reports of increased infection rates abd describing the revision of the product labeling with additional statements that relate to product usage and reinforcement of standard clinical practices and proper aseptic technique. The revised labeling statements are as follows: * Ensure clamp is open. Swab Luer activated surface with preferred antiseptic prior to first use and before every subsequent connection. * Don not swab Luer activated surface when clamp is closed or valve is recessed. Ineffective swabbibg may result. Replace if valve remains recessed. * Flush Luer activated valve after blood infusion/sampling. If valve cannot be cleared of blood, replace immediately. * Luer lock vacutainer holders, syringes and accessories are recommended for blood sampling. If blood remains on valve surface after sampling, swab surface. * If intermittently disconnecting set from Luer activated valve, immediately cover male Luer with a sterile replacement protector.
- Quantity
- 29,425,708 sets
- Distribution
- Nationwide, and internationally through Baxter subsidiaries to Macedonia, Nigeria, Saudi Arabia, South Africa, Ireland, Italy, Denmark, Germany, the Netherlands, Hong Kong, Indonesia, Singapore, Spain, Belgium, Norway, United Kingdom, France and Luxembourg.
- Lot, serial or model codes
- all lots of product code 2N8399, 1C8678, 1C8680, 1C8683, 1C8684, 1C8687, 1C8693, 1C8702, 1C8704, 1C8705, 1C8706, 1C8707, 1C8712, 1C8722, 1C8723, 1C8727, 2C0168, 2C6254, 2C6255, 2C6256, 2C7461, 2C7462, 2C8401, 2C8402, 2C8419, 2C8425, 2C8428, 2C8515, 2C8519, 2C8537, 2C8541, 2C8546, 2C8548, 2C8571, 2C8593, 2C8606, 2C8607, 2C8610, 2C8612, 2C8632, 2C8634, 2C8671, 2C8700, 2C8720, 2C8750, 2C8751, 2C8819, 2C8851, 2C8857, 2C8858, 2C8860, 2C8862, 2C8864, 2C8865, 2C8875, 2C8891, 2C8895, 2C8920, 2C8921, 2C8930, 2C8931, 2H7462, 2H7463, 2H8401, 2H8480, 2H8486, 2H8519, 2H8603, 2H8720, 2H8750, 2H8819, 2H8860, 2H8862, 2H8864, 2H8865, 2N8300, 2N8334, 2N8335, 2N8341, 2N8371, 2N8373, 2N8374, 2N8377, 2N8378, 2N9204K, 2N9205K, 3C0122, 3C0134, 3C0139, 3C0143, 3C0148, 3C0150, 3C0151, 3C0157, 3C0158, 3C0159, 3C0160, 3C0162, 3C0165, 4C8714, 4C8723, 4H8723
- 510(k) numbers
- ["K832284","K981792","K984381","K993120","K003225"]
- PMA numbers
- not on file
- Product code
- FPA
- Firm FEI
- 1416980
- Firm city
- Round Lake
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28