Synthes (USA) Products LLC: Synthes Threaded Plate Holder- Long for 3.5mm Locking Hold; Part Numbers 324.031. The Threaded Plate Holder can be used as an aid to position the plate on the bone. The plate holder may also function as an insertion handle for use with minimally invasive plating techniques. The plate may be temporarily held in place with standard plate holding forceps or the push-pull reduction device.
- Recall number
- Z-0192-2016
- Recalling firm
- Synthes (USA) Products LLC
- Product
- Synthes Threaded Plate Holder- Long for 3.5mm Locking Hold; Part Numbers 324.031. The Threaded Plate Holder can be used as an aid to position the plate on the bone. The plate holder may also function as an insertion handle for use with minimally invasive plating techniques. The plate may be temporarily held in place with standard plate holding forceps or the push-pull reduction device.
- Status
- Terminated
- Initiated
- 2015-10-07
- Posted
- not on file
- Terminated
- 2016-08-26
- Reason
- The affected part number and lot listed above is incorrectly etched with the wrong part number and may not fit into a Small Fragment Plate.
- Root cause
- Mixed-up of materials/components
- Action
- The firm, DePuy Synthes, sent an "URGENT NOTICE: MEDICAL DEVICE RECALL" letter dated 10/5/2015 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: " Call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number and replacement product(s). " Complete the Verification Section (page 3 of this letter) by checking the appropriate box indicating affected product has been located. Also, please indicate the number of devices found and note the Return Authorization Number. Please include your name, title, address, telephone number and signature in the spaces provided. " Return the Verification Section (page 3 of this letter) with the product to: Credit/Returns, DePuy Synthes, 1101 Synthes Avenue, Monument, CO 80132. Even if you do not have any product. " Send a copy of the completed Verification Section by: Fax: 844-721-3048 or Scan/email: Synthes4704@stericycle.com. " Note: If the Verification Section is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual on page 3 of the notification. If you have any questions, please call 610-719-5450 or contact your DePuy Synthes Sales Consultant.
- Quantity
- 25
- Distribution
- Worldwide distribution: US (nationwide) and country of: Canada.
- Lot, serial or model codes
- Threaded Plate Holder-Long for 3.5mm Locking Hole Part Numbers Lot Numbers 324.031 8225613
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- EZX
- Firm FEI
- 3008812560
- Firm city
- West Chester
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28