FDA Medical Device Recalls (openFDA)
Sorin Group USA, Inc.: Revolution Blood Pump, "revOlution Centrifugal pump Catalog# 050300000***Rx ONLY***SORIN GROUP ITALIA via Statale 12 Nord, 86 41037 MIRANDOLA (MO) (Italy) Distributed in the U.S. by: Sorin Group USA, Inc. 14401 W. 65th way Arvada (CO) 80004-3599" Intended for use in cardiopulmonary bypass procedures.
- Recall number
- Z-0192-2010
- Recalling firm
- Sorin Group USA, Inc.
- Product
- Revolution Blood Pump, "revOlution Centrifugal pump Catalog# 050300000***Rx ONLY***SORIN GROUP ITALIA via Statale 12 Nord, 86 41037 MIRANDOLA (MO) (Italy) Distributed in the U.S. by: Sorin Group USA, Inc. 14401 W. 65th way Arvada (CO) 80004-3599" Intended for use in cardiopulmonary bypass procedures.
- Status
- Terminated
- Initiated
- 2009-09-25
- Posted
- 2009-11-13
- Terminated
- 2009-12-14
- Reason
- A weakness in the plastic packaging tray could cause the tray to crack; compromising product sterility.
- Root cause
- Packaging change control
- Action
- Customers were notified by an Urgent: Field Safety Notice, dated 9/25/09, and instructed to: check their inventory for the affected product, complete the customer response form and return or visually inspect all affected product before use. For additional information contact Sorin Group Customer Service at 1-800-650-2623.
- Quantity
- 808 units
- Distribution
- Worldwide Distribution -- US, Canada, Japan, and Europe.
- Lot, serial or model codes
- Lot numbers: 0905220163, 0905220166, 0906240182, and 0906240185.
- 510(k) numbers
- ["K011835"]
- PMA numbers
- not on file
- Product code
- KFM
- Firm FEI
- 1718850
- Firm city
- Arvada
- Firm state
- CO
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as