Skip to the content

FDA Medical Device Recalls (openFDA)

Sorin Group USA, Inc.: Revolution Blood Pump, "revOlution Centrifugal pump Catalog# 050300000***Rx ONLY***SORIN GROUP ITALIA via Statale 12 Nord, 86 41037 MIRANDOLA (MO) (Italy) Distributed in the U.S. by: Sorin Group USA, Inc. 14401 W. 65th way Arvada (CO) 80004-3599" Intended for use in cardiopulmonary bypass procedures.

Recall number
Z-0192-2010
Recalling firm
Sorin Group USA, Inc.
Product
Revolution Blood Pump, "revOlution Centrifugal pump Catalog# 050300000***Rx ONLY***SORIN GROUP ITALIA via Statale 12 Nord, 86 41037 MIRANDOLA (MO) (Italy) Distributed in the U.S. by: Sorin Group USA, Inc. 14401 W. 65th way Arvada (CO) 80004-3599" Intended for use in cardiopulmonary bypass procedures.
Status
Terminated
Initiated
2009-09-25
Posted
2009-11-13
Terminated
2009-12-14
Reason
A weakness in the plastic packaging tray could cause the tray to crack; compromising product sterility.
Root cause
Packaging change control
Action
Customers were notified by an Urgent: Field Safety Notice, dated 9/25/09, and instructed to: check their inventory for the affected product, complete the customer response form and return or visually inspect all affected product before use. For additional information contact Sorin Group Customer Service at 1-800-650-2623.
Quantity
808 units
Distribution
Worldwide Distribution -- US, Canada, Japan, and Europe.
Lot, serial or model codes
Lot numbers: 0905220163, 0905220166, 0906240182, and 0906240185.
510(k) numbers
["K011835"]
PMA numbers
not on file
Product code
KFM
Firm FEI
1718850
Firm city
Arvada
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register