CSL Behring GmbH: SIEMENS N/T Protein Control SL for use on the BN II, BN ProSpec Systems: N/T Protein Control SL/L: REF/Product Code OQIN19, SMN 10446079 (Low); N/T Protein Control SL/M, REF/Product Code OQIO19, SMN 10446085 (Mid); N/T Protein Control SL/H, REF/Product Code OQIP19, SMN 10446089 (High);

Recall number
Z-0191-2018
Recalling firm
CSL Behring GmbH
Product
SIEMENS N/T Protein Control SL for use on the BN II, BN ProSpec Systems: N/T Protein Control SL/L: REF/Product Code OQIN19, SMN 10446079 (Low); N/T Protein Control SL/M, REF/Product Code OQIO19, SMN 10446085 (Mid); N/T Protein Control SL/H, REF/Product Code OQIP19, SMN 10446089 (High);
Status
Terminated
Initiated
2017-06-05
Posted
not on file
Terminated
2020-04-07
Reason
Siemens Healthcare Diagnostics has confirmed a negative bias of approximately -20% for results obtained with N Antiserum to Human Ceruloplasmin when calibrating Ceruloplasmin with various lots of N Protein Standard SL. N/T Protein Controls SL for Ceruloplasmin are similarly biased. --- The affected standards and controls for the affected lots were incorrectly value assigned, which led to a lower result compared to correctly assigned standards. Note: N Antiserum to Human Ceruloplasmin (REF OUIE) works as intended. --- Siemens investigation indicated that N Protein Standard SL and corresponding N/T Protein Control SL (Levels Low, Mid and High) are inaccurately assigned for Ceruloplasmin.
Root cause
Nonconforming Material/Component
Action
URGENT MEDICAL DEVICE CORRECTION Letter (PP17-011.a.US, dated May 30, 2017) was sent to US consignees via FedEx on 6/5/2017. URGENT FIELD CORRECTIVE ACTION Letter was e-mailed to customers Outside the US. Customers are instructed to use the re-assigned Ceruloplasmin values for the lots indicated in the letter.
Quantity
not on file
Distribution
Worldwide including the countries of United States, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia-Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Cyprus, Denmark, Egypt, Finland, France, Germany, Greece, Great Britain, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, japan, Kazakhstan, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, Mexico, Morocco, Netherlands, New Zealand, Norway, Poland, Portugal, Republic of Korea (S), Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, South Africa, Switzerland, Syria, Taiwan, Thailand, Trinidad-Tobago, Turkey, Turkmenistan, United Arab Emirates, Uruguay, Vatican, Venezuela,
Lot, serial or model codes
SMN 10446079: Lot (UDI): 084647A (00842768016127084647A20170913), 084647C (00842768016127084647C20170913) exp. date 2017-09-13; 084648(0084276801612708464820180430),  084648C (00842768016127084648C20180430),  084648H (00842768016127084648H20180430), exp. date 2018-04-30. --- SMN 10446085: Lot (UDI): 084749D (00842768016134084749D20170917) exp. date 2017-09-17; 084750 (03 0084276801613408475020180203),  084750B (00842768016134084750B20180203), 084750E (00842768016134084750E20180203) exp. date 2018-02-03. --- SMN 10446089: Lot (UDI) 084847A (00842768016110084847A20170613),  084847B (00842768016110084847B20170613) exp. date 2017-06-13; 084848A (00842768016110084848A20171217),  084848D (00842768016110084848D20171217),  084848E (00842768016110084848E20171217),  084848H (00842768016110084848H20171217),  084848L (00842768016110084848L20171217) exp. date 2017-12-17;  084849 (0084276801611008484920180704),  084849C (00842768016110084849C20180704), 084849D (00842768016110084849D20180704) 084849E (00842768016110084849E20180704) exp. date 2018-07-04.
510(k) numbers
["K052788"]
PMA numbers
not on file
Product code
CHN
Firm FEI
3003098680
Firm city
Marburg
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register