FDA Medical Device Recalls (openFDA)
Allez Spine, LLC: Size 8 Pedicle Screws of the Allez Spine Laguna Pedicle Screw System, Screw Size (diameter x length) 8 x 80, Catalog Number LG-PR0880, Allez Spine, LLC., Irvine, CA
- Recall number
- Z-0191-2008
- Recalling firm
- Allez Spine, LLC
- Product
- Size 8 Pedicle Screws of the Allez Spine Laguna Pedicle Screw System, Screw Size (diameter x length) 8 x 80, Catalog Number LG-PR0880, Allez Spine, LLC., Irvine, CA
- Status
- Terminated
- Initiated
- 2007-07-30
- Posted
- 2008-03-11
- Terminated
- 2012-04-09
- Reason
- Screw Shank Failure; separation of the screw shank portion from the pedicle head portion of the Size 8 Pedicle Screw, when torquing down the single piece locking nut during implantation.
- Root cause
- Device Design
- Action
- On July 30, 2007 Allez Spine notified all its field representatives and distributors via email of a potential for failure of the Size 8 Pedicle Screws when used in combination with the single piece locking nuts and instructed them to return all unused inventory of Size 8 Screws and the two piece Locking Nuts with saddle. This action was to be carried out to the retail level-hospital, clinic, lab, and physicians sold the product. The email also had an attachment of a list of expected frequently asked questions (FAQ) from surgeons and answers to help them respond. A second Notice to all users was sent by the recalling firm on Feb 29, 2008, to request follow-up of implanted units.
- Quantity
- 14
- Distribution
- Nationwide: including the states of AZ, CA, CO, GA, NV, TX and UT
- Lot, serial or model codes
- Lot Number: 10528
- 510(k) numbers
- ["K050060"]
- PMA numbers
- not on file
- Product code
- MNI
- Firm FEI
- 3005032381
- Firm city
- Irvine
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as