Skip to the content

FDA Medical Device Recalls (openFDA)

Allez Spine, LLC: Size 8 Pedicle Screws of the Allez Spine Laguna Pedicle Screw System, Screw Size (diameter x length) 8 x 80, Catalog Number LG-PR0880, Allez Spine, LLC., Irvine, CA

Recall number
Z-0191-2008
Recalling firm
Allez Spine, LLC
Product
Size 8 Pedicle Screws of the Allez Spine Laguna Pedicle Screw System, Screw Size (diameter x length) 8 x 80, Catalog Number LG-PR0880, Allez Spine, LLC., Irvine, CA
Status
Terminated
Initiated
2007-07-30
Posted
2008-03-11
Terminated
2012-04-09
Reason
Screw Shank Failure; separation of the screw shank portion from the pedicle head portion of the Size 8 Pedicle Screw, when torquing down the single piece locking nut during implantation.
Root cause
Device Design
Action
On July 30, 2007 Allez Spine notified all its field representatives and distributors via email of a potential for failure of the Size 8 Pedicle Screws when used in combination with the single piece locking nuts and instructed them to return all unused inventory of Size 8 Screws and the two piece Locking Nuts with saddle. This action was to be carried out to the retail level-hospital, clinic, lab, and physicians sold the product. The email also had an attachment of a list of expected frequently asked questions (FAQ) from surgeons and answers to help them respond. A second Notice to all users was sent by the recalling firm on Feb 29, 2008, to request follow-up of implanted units.
Quantity
14
Distribution
Nationwide: including the states of AZ, CA, CO, GA, NV, TX and UT
Lot, serial or model codes
Lot Number: 10528
510(k) numbers
["K050060"]
PMA numbers
not on file
Product code
MNI
Firm FEI
3005032381
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register