FDA Medical Device Recalls (openFDA)
Baxter Healthcare Corp.: SIGMA SPECTRUM Infusion Pump with Master Drug Library component (SHARP LCD screen, 55074-2 revision B) Product Usage: Intended to be used as a visual aid during the controlled administration of intravenous fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy.
- Recall number
- Z-0189-2014
- Recalling firm
- Baxter Healthcare Corp.
- Product
- SIGMA SPECTRUM Infusion Pump with Master Drug Library component (SHARP LCD screen, 55074-2 revision B) Product Usage: Intended to be used as a visual aid during the controlled administration of intravenous fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy.
- Status
- Terminated
- Initiated
- 2013-09-24
- Posted
- 2013-11-08
- Terminated
- 2016-11-09
- Reason
- Baxter Healthcare Corporation is recalling the LCD display screen on the SIGMA SPECTRUM Infusion Pump with Master Drug Library for not meeting standards for withstanding an electrostatic discharge.
- Root cause
- Under Investigation by firm
- Action
- Baxter Healthcare Corporation sent an Urgent Product Recall letter dated September 24, 2013 to all affected customers via first class mail. The letter identified the affected product, problem and actions to be taken. Customers were instructed to locate the affected product, remove them from use and acknowledge your receipt of the recall notification by completing the provided Customer Reply Form and fax to 224-270-5467. Customers were asked to contact Baxter Healthcare Medina at 800-356-3454 (choose option 1) Monday through Friday during the hourse of 7:00 am and 7:00 pm Eastern Time to order replacement device(s). For technical questions contact Baxter Healthcare Medina at 800-356-3454, (choose option1) and for clinical questions contact Medical Information Services at Baxter 800-933-0303.
- Quantity
- 346 units
- Distribution
- Worldwide Distribution - USA Nationwide in the states of: AR, AZ, CA, CO, CT, FL, HI, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NM, NY, OH, OK, OR, PA, TN, UT, VA, WA and WV and the country Canada.
- Lot, serial or model codes
- Product Code: 35700BAX & 35700ABB; Serial Numbers: 433713, 435563, 704056, 707290, 715315, 724216, 740127, 752318, 752447, 753675, 754947, 759766, 765934, 770034, 772017, 773470, 776507, 781016, 781969, 783352, 783454, 787567, 789720, 791225, 794828, 794990, 797185, 804798, 808393, 808764, 810971, 812076, 818718, 821213, 824067, 824198, 839143, 850028, 850577, 852007, 852431, 855450, 860201, 864799, 876765, 877548, 877944, 878637, 879810, 879886, 880022, 884707, 884778, 885565, 889416, 889814, 890949, 890987, 894957, 895413, 896662, 896912, 897576, 900401, 904778, 913511, 915035, 915981, 916173, 916948, 919159, 921162, 921937, 922515, 923861, 945583, 953011, 953360, 953995, 955881, 959847, 967694, 978892, 981692, 986022, 997333, 998512, 1002325, 1018432, 1018436, 1018737
- 510(k) numbers
- ["K042121"]
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 1417572
- Firm city
- Deerfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as