Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Renal Div: HomeChoice PRO and Yume Plus Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C8310, 5C8310R, T5C8300, T5C8300R

Recall number
Z-0189-05
Recalling firm
Baxter Healthcare Renal Div
Product
HomeChoice PRO and Yume Plus Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C8310, 5C8310R, T5C8300, T5C8300R
Status
Terminated
Initiated
2004-11-01
Posted
2004-11-11
Terminated
2007-12-11
Reason
The on/off power switch is not grounded as a separate component in the device, and can loosen from its housing, which may result in an electrical shock to the user.
Root cause
Other
Action
Baxter sent Urgent Device Correction letters to the Peritoneal Dialysis Home Care Coordinators on 11/1/04 and to the Home Dialysis Patients on 11/2/04. The consignees were informed of the potential for a shock if the power switch loosens in the housing, and that the switch would be replaced with a grounded power switch on a next service call basis. If the patient experiences any difficulty turing the device on or off, or if they notice that the power switch is loose, they were instructed to immediately unplug the device and contact Baxter Global Technical Service (BGTS) at 1-800-553-6898, option 1, for immediate assistance.
Quantity
10,800 units
Distribution
Nationwide, including Puerto Rico, and internationally to Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Ecuador, Guatemala, Honduras, Hong Kong, India, Japan, Korea, Malaysia, Nicaragua, New Zealand, Peru, Philippines, Saudi Arabia, Singapore, South Africa, Taiwan, Turkey, Uruguay and Venezuela.
Lot, serial or model codes
All units that do not bear the CE marking.
510(k) numbers
["K012988"]
PMA numbers
not on file
Product code
FKX
Firm FEI
3000210122
Firm city
Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register