Medtronic Navigation, Inc.: StealthStation S8 Application version 2.0 and 2.0.1 (Part 9735762) The StealthStation System, with StealthStation Cranial software, is intended as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures. The StealthStation FlexENT System, with StealthStation ENT software, is intended as an aid for locating anatomical structures in either open or percutaneous ENT procedures.

Recall number
Z-0188-2024
Recalling firm
Medtronic Navigation, Inc.
Product
StealthStation S8 Application version 2.0 and 2.0.1 (Part 9735762) The StealthStation System, with StealthStation Cranial software, is intended as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures. The StealthStation FlexENT System, with StealthStation ENT software, is intended as an aid for locating anatomical structures in either open or percutaneous ENT procedures.
Status
Open, Classified
Initiated
2023-09-27
Posted
2023-11-10
Terminated
not on file
Reason
Due to a software anomaly which potentially could result in the surgical planning data shifting to an unintended location.
Root cause
Software design
Action
On 09/27/2023, the firm sent an "URGENT: MEDICAL DEVICE CORRETION" Letter/Communication via UPS to customer informing them of a software anomaly affecting certain StealthStation S8 and StealthStation FlexENT system with the StealthStation S8 App software version 2.0 and 2.0.1. Due to the software anomaly, there is the potential for the surgical planning data to encounter a displayed shift which could result in proceeding to an incorrect location from the planned target. This result in unintended tissue damage leading to permanent neurological injury, additional pass of device (biopsy needle or electrode), prolonged procedure or need for additional surgery. Customer are instructed to: 1. Follow the below instructions and refer to Appendix A for full details on impacted systems, issues and mitigations: a. Do not change the reference exam in Cranial or ENT procedures, if the surgical planning data is defined on a reference exam merged with pre-merge type or diffusion series. 2. Please review this information, including the additional details in Appendix A, with all physician users. If you have any questions related to this issue, please contact Medtronic Technical Services for help at 1-888-826-5603 or email at rs.navtechsupport@medtronic.com 3. Please confirm via the enclosed confirmation form that this notification has been communicated within your facility with all physician users. Send the completed Customer Confirmation Form to Medtronic via email at neuro.quality@medtronic.com. a. Medtronic will contact your facility to coordinate the software update when it is available. 4. This notice needs to be passed on to those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. Questions or assistance, contact Medtronic Sales Representatives or Technical Services at 1-888-826-5603.
Quantity
5,240 affected systems
Distribution
Worldwide Distribution: U.S. (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. O.U.S.(foreign) including countries of: ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BRAZIL, BULGARIA, CANADA, CANARY ISLANDS, CHILE, COLOMBIA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, FRENCH POLYNES, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KAZAKHSTAN, KENYA, KOREA, KUWAIT, LATVIA, LITHUANIA, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PANAMA, PHILIPPINES, POLAND, PORTUGAL, REUNION, ROMANIA, RUSSIAN FEDERATION, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UKRAINE, and UNITED KINGDOM
Lot, serial or model codes
Model Number: 9735762 UDI-DI Code: 00763000532123 Software Versions 2.0 and 2.0.1 All systems with the impacted software version regardless of how it was purchased Updated list of Product Models/Serial provide on 06/06/2024 and was uploaded to Associated Documents.
510(k) numbers
["K212397"]
PMA numbers
not on file
Product code
HAW
Firm FEI
1000517638
Firm city
Lafayette
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register