Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Molecular Systems, Inc.: cobas 4800 BRAF V600 Mutation Test for In Vitro Diagnostic Use Roche Molecular Systems, Inc., Product Usage: The primary use of the cobas 4800 BRAF V600 Mutation Test is the detection of the BRAF V600 mutations in DNA extracted from formalin-fixed, paraffin-embedded human melanoma tissue. It is intended to be used as an aid in selecting melanoma patients whose tumors carry BRAF V600 mutations for treatment with Zelboraf (vemurafenib).

Recall number
Z-0188-2013
Recalling firm
Roche Molecular Systems, Inc.
Product
cobas 4800 BRAF V600 Mutation Test for In Vitro Diagnostic Use Roche Molecular Systems, Inc., Product Usage: The primary use of the cobas 4800 BRAF V600 Mutation Test is the detection of the BRAF V600 mutations in DNA extracted from formalin-fixed, paraffin-embedded human melanoma tissue. It is intended to be used as an aid in selecting melanoma patients whose tumors carry BRAF V600 mutations for treatment with Zelboraf (vemurafenib).
Status
Terminated
Initiated
2012-07-31
Posted
2012-11-02
Terminated
2013-09-11
Reason
It was identified that BRAF controls dispositioned as "Internal Use Only" were incorrectly packaged into cobas 4800 BRAF V600 Mutation Test CE-IVD kit lot R03607 and released for distribution. Two control batches manufactured during two process validation studies are involved: PV7711: BRAF Mutant (MUT) Control PV0853: BRAF Wild Type (WT) Control
Root cause
Mixed-up of materials/components
Action
Roche sent a Safety Board Notice - Product Bulletin 2012-013 on 31-Jul-2012 to affected customers. The notice identifed the affected product, description of situation and actios to be taken. Customers were instructed to discontinue use, quarantine and discard any remaining kits of the affected product in inventory. Customers were advised to contact their affiliate sales office for replacement. The notice states a new cobas 4800 BRAF V600 Mutation Test lot (R07749) is anticipated to be available on 03-Aug-2012. For questions call +01.908.253.7569.
Quantity
44 kits
Distribution
Worldwide Distribution - Australia, Austria, Belgium, China, Hungary, Turkey and United Kingdom.
Lot, serial or model codes
Material number 05985595190; Lot number R03607
510(k) numbers
not on file
PMA numbers
["P110020"]
Product code
OWD
Firm FEI
2243471
Firm city
S Branchburg
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register