FDA Medical Device Recalls (openFDA)

Alere San Diego: PDI Alcohol Prep Pads packaged in the INRatio kit. PDI Alcohol Prep Pads, 100/box. Alcohol Prep Pads, packaged in the Alere INRatio 2 PT/INR Home Monitoring Kit; or Hemosense INRatio 2 Prothrombin Time/INR Testing Kit, Self-Test System; Alcohol Prep Pads, packaged in the Hemosense INRatio Prothrombin Time/INR Testing Kit, Self-Test System; Alcohol Prep Pads, packaged in the INRatio Starter Kit Product Model Number: 0200086, 0200432, 0100007, 0100072 The Alere INRatio2 PT/INR Home Monitoring System is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The Alere INRatio2 PT/INR Home Monitoring System is intended for use outside the body (in vitro diagnostic use) by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The Alere INRatio2 PT/ INR Home Monitoring System is not intended to be used for screening purposes. The Alere INRatio2 PT/INR Home Monitoring System should only be used with a prescription from your physician. The INRatio2 PT Monitoring system is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The INRatio2 PT Monitoring system is intended for use outside the body (in vitro diagnostic use) by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The INRatio2 PT Monitoring System is not intended to be used for screening purposes. The INRatio2 PT Monitoring system should only be used with a prescription from your physician. The INRatio system performs Prothrombin Time (PT) testing on fresh, capillary whole blood. The PT test is the principle assay used to monitor warfarin therapy. Patients taking warfarin and other oral anticoagulants should consult with their healthcare provider before adjusting their dosage. The thromboplastins used in performing PT tests can vary substantially causing differences in results between methods. The International Normalized Ratio (INR) is a mathematical correction of the PT result that adjusts for differences in thromboplastins. Consult your physician for the appropriate therapeutic range for you.

Recall number
Z-0188-2012
Recalling firm
Alere San Diego
Product
PDI Alcohol Prep Pads packaged in the INRatio kit. PDI Alcohol Prep Pads, 100/box. Alcohol Prep Pads, packaged in the Alere INRatio 2 PT/INR Home Monitoring Kit; or Hemosense INRatio 2 Prothrombin Time/INR Testing Kit, Self-Test System; Alcohol Prep Pads, packaged in the Hemosense INRatio Prothrombin Time/INR Testing Kit, Self-Test System; Alcohol Prep Pads, packaged in the INRatio Starter Kit Product Model Number: 0200086, 0200432, 0100007, 0100072 The Alere INRatio2 PT/INR Home Monitoring System is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The Alere INRatio2 PT/INR Home Monitoring System is intended for use outside the body (in vitro diagnostic use) by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The Alere INRatio2 PT/ INR Home Monitoring System is not intended to be used for screening purposes. The Alere INRatio2 PT/INR Home Monitoring System should only be used with a prescription from your physician. The INRatio2 PT Monitoring system is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The INRatio2 PT Monitoring system is intended for use outside the body (in vitro diagnostic use) by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The INRatio2 PT Monitoring System is not intended to be used for screening purposes. The INRatio2 PT Monitoring system should only be used with a prescription from your physician. The INRatio system performs Prothrombin Time (PT) testing on fresh, capillary whole blood. The PT test is the principle assay used to monitor warfarin therapy. Patients taking warfarin and other oral anticoagulants should consult with their healthcare provider before adjusting their dosage. The thromboplastins used in performing PT tests can vary substantially causing differences in results between methods. The International Normalized Ratio (INR) is a mathematical correction of the PT result that adjusts for differences in thromboplastins. Consult your physician for the appropriate therapeutic range for you.
Status
Terminated
Initiated
2011-10-11
Posted
2011-11-16
Terminated
2012-02-29
Reason
Alere has initiated a recall of the Professional Disposables International (PDI) alcohol prep products due to the potential presence of low levels of the microorganism Bacillus cereus.
Root cause
Material/Component Contamination
Action
Alere sent a recall notification letter dated October 6, 2011, and Verification Form to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to use an alternate prep pad that is not involved with the PDI recall, or use a sterile gauze pad with isopropyl alcohol. If customers purchase alcohol pads from a store or pharmacy, customers can confirm with their pharmacist that the pads they are purchasing are not associated with the PDI recall. INRatio/INRatio2 kits at Alere are currently being reworked to include a sterile PDI alcohol pad. Customers were instructed to complete and FAX the enclosed Verification Form within 10 days to confirm their receipt of the notice and to indicate the number of discarded/required replacement alcohol prep pads. Customers with questions about INRatio/ INRatio 2 were instructed to consult their healthcare provider, or call Alere Technical Service at 877-866-5313 (24 hours a day, 7 days a week).
Quantity
Estimated 14,248 Total units
Distribution
Nationwide Distribution
Lot, serial or model codes
Lot/Unit Codes: 245765R, 245763R, 245766R, 245768, 245769, 252549, 245770, 252020, 252021, 252530, 252532, 252533, 256080, 256081, 256082, 256083, 256084, 256085, 260082, 256086, 256087, 256088, 256089, 256090, 256091, 256092, 261879, 261880.
510(k) numbers
["K020679","K021923","K072727"]
PMA numbers
not on file
Product code
LKB
Firm FEI
1000125596
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register