Skip to the content

FDA Medical Device Recalls (openFDA)

Physio-Control, Inc.: LIFEPAK(R) Electrode Starter Kits, Catalog # 11101-000017, MIN #3202784-009 The defibrillation electrodes are used only with LIFEPAK EXPRESS(R) AED, LIFEPAK CR(R) Plus AED, LIFEPAK 1000 defibrillator, or LIFEPAK 500 Biphasic AED s with a pink connector.

Recall number
Z-0187-2018
Recalling firm
Physio-Control, Inc.
Product
LIFEPAK(R) Electrode Starter Kits, Catalog # 11101-000017, MIN #3202784-009 The defibrillation electrodes are used only with LIFEPAK EXPRESS(R) AED, LIFEPAK CR(R) Plus AED, LIFEPAK 1000 defibrillator, or LIFEPAK 500 Biphasic AED s with a pink connector.
Status
Terminated
Initiated
2017-10-27
Posted
2017-12-12
Terminated
2019-06-12
Reason
The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols.
Root cause
Process control
Action
The firm Physio Control, initiated their field action on 10/27/2017 by issuing a press release. The firm followed this with a recall notice dated 10/23/2017, to the direct consignees on 11/03/2017. The letter described the product, explained the problem and provided the following instructions: Quarantine the affected lots;contact Physio-Control at 1-866-231-1220, 6:00 A.M. to 4:00 P.M. (Pacific), Monday Friday to arrange for replacement of your unused electrodes and "Upon receipt of this notification, post a copy of the enclosed Correct Electrode and Packaging Labeling with each of your AEDs, which shows the correct placement of the electrodes." Physio_Control will provide replacement products for all unused affected Electrodes. Should you have any questions about this Product Recall, please contact us at 1-866-231-1220, 6:00 A.M. to 4:00 P.M. (Pacific), Monday Friday.
Quantity
14,219 units in total
Distribution
Worldwide distribution: US (nationwide) including Puerto Rico and Guam and countries of: Canada, Austria, Belgium, Czech Republic; Denmark, Finland, France, French Polynesia, Germany, Hungary, Ireland, Italy, Latvia, Netherlands, Australia, China, Hong Kong, Japan, Argentina, Mexico, Norway, Panama, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom, and New Zealand.
Lot, serial or model codes
45932237, 46042286, 45979590, 46050960, 45979954, 46052545, 46007867, 46061770, 46023185, 46063054, 46023823, 46078012,
510(k) numbers
["K022732"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
3015876
Firm city
Redmond
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register