FDA Medical Device Recalls (openFDA)
Invacare Corporation: Invacare 600 lb. Bariatric Bed, Head motor extension cable used on component numbers BAR5490 and BAR5490IVC. These components are part of the following Bed Model Number(s): BAR600, BARPKGCA, BAR600IVC, BARPKGIVC-1633, and BARPKG.
- Recall number
- Z-0187-2012
- Recalling firm
- Invacare Corporation
- Product
- Invacare 600 lb. Bariatric Bed, Head motor extension cable used on component numbers BAR5490 and BAR5490IVC. These components are part of the following Bed Model Number(s): BAR600, BARPKGCA, BAR600IVC, BARPKGIVC-1633, and BARPKG.
- Status
- Terminated
- Initiated
- 2011-06-23
- Posted
- 2011-11-14
- Terminated
- 2013-07-02
- Reason
- Invacare Corporation decided to recall the products because risk analysis reveals that miskeying or disconnection of the head motor extension cable during bed set-up may result in product damage, or in rare instances, the potential for fire.
- Root cause
- Labeling design
- Action
- On 9/19/11, Invacare sent Recall Notificaiton Letters to their customers. They instructed their customers to quarantine the product, complete and return a response form.
- Quantity
- 35,747
- Distribution
- Distribution Nationwide and to United Arab Emirates, Dominican Republic, Egypt, Guam, Mexico, & Virgin Islands.
- Lot, serial or model codes
- Model Number(s): BAR600, BARPKGCA, BAR600IVC, BARPKGIVC-1633, and BARPKG. 35747 serial numbers available from Invacare.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LLI
- Firm FEI
- 1525712
- Firm city
- Elyria
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as