Skip to the content

FDA Medical Device Recalls (openFDA)

CareFusion 303, Inc.: BD Alaris Syringe Module, REF 8110

Recall number
Z-0185-2024
Recalling firm
CareFusion 303, Inc.
Product
BD Alaris Syringe Module, REF 8110
Status
Open, Classified
Initiated
2023-09-15
Posted
2023-11-07
Terminated
not on file
Reason
Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Root cause
Under Investigation by firm
Action
On September 15, 2023, CareFusion a subsidiary of Becton Dickinson (BD) posted an "Urgent Medical Device Labeling Correction" on their website. On October 4, 2023 CareFusion mailed a copy of the correction notification via FedEx. CareFusion asked consignees to take the following actions: 1. You should cease use of Cardinal Health branded Monoject" syringes with the BD Alaris" Syringe and PCA Modules; except for SKU 8881135609, which is not impacted by this notification. 2. Please refer to Attachments 1 and 2 of this notice for a full selection of syringes that are compatible with the BD Alaris" Syringe and PCA Modules. 3. Replace your current BD Alaris" Syringe and PCA Module Compatibility Lists with Attachments 1 and 2 of this notice. 4. Please circulate this notice to all those who require awareness within your organization or to any organization whom you may have transferred BD Alaris" Infusion System devices to. 5. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification. 6. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787
Quantity
133,727 units
Distribution
Worldwide Distribution: US (nationwide): AK, AL, AR, AS, AZ, CA, CO, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; OUS (foreign): Canada, Germany, Cuba, Italy, Japan, South Korea, Turkey, United Arab Emirates, Australia, Bahrain, Botswana, Kuwait, Mexico, Netherlands, Philippines, Pakistan, Qatar, South Africa, Singapore, Belgium, Israel, India, New Zealand, Malaysia, Taiwan, Switzerland
Lot, serial or model codes
All Lots/ UDI-DI:10885403516047, 10885403515323, 10885403811043, 10885403515255 10885403811036, 10885403515262, 10885403811012, 10885403515279, 10885403424267
510(k) numbers
["K051641","K072105","K133532","K211218"]
PMA numbers
not on file
Product code
FRN
Firm FEI
2016493
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register