Implant Direct Sybron Manufacturing, LLC: ImplantDirect Legacy(TM) 3 Implant, 13MM L, REF 864713 This Legacy implant is a two-piece implant for one-stage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework.

Recall number
Z-0185-2018
Recalling firm
Implant Direct Sybron Manufacturing, LLC
Product
ImplantDirect Legacy(TM) 3 Implant, 13MM L, REF 864713 This Legacy implant is a two-piece implant for one-stage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework.
Status
Terminated
Initiated
2017-10-10
Posted
not on file
Terminated
2018-09-27
Reason
One reported complaint and discovered that the outer vial was out of round and undersized. Therefore a portion of the opening may not allow inner vial and components to slide freely out of the vial. As a result, the cap could possibly not be screwed on properly and the sterilization compromised.
Root cause
Process design
Action
The firm, Implant Direct Sybron Manufacturing, LLC, sent an "URGENT: Safety Alert/Advisory Notice" letter to its distributors/customers on 10/10/2017. Distributors were instructed as follows: "1. Please review your inventory for the affected product. 2. Please complete and return the Acknowledgement and Recall Return Form within 48 hours. 3. If you are an authorized Implant Direct Sybron Manufacturing distributor, we request that you identify those customers that may have been shipped the affected product lot and contact these customers to inform them of this issue within fortyeight (48) hours of receipt of this notification in order to provide the customers with replacement product." Enclosed within this notification, you will find a padded return envelope with replacement product. Please return affected product in the padded envelope. If you have any questions contact Implant Direct Sybron Manufacturing LLC Customer Care at 1-888-649-6425. Customers were instructed as follows: "1. Please review your inventory for the affected product. 2. Please complete and return the Acknowledgement and Recall Return Form within 48 hours. Please share this information with your staff and retain this notification as part of your documentation. Enclosed within this notification, you will find a padded return envelope with replacement product. Please return affected product in the padded envelope. If you have any questions contact Implant Direct Sybron Manufacturing LLC Customer Care at 1-888-649-6425."
Quantity
75 units
Distribution
Worldwide distribution: US (nationwide) including Puerto Rico and country of Japan.
Lot, serial or model codes
Lot Number: 79607
510(k) numbers
["K090234"]
PMA numbers
not on file
Product code
DZE
Firm FEI
3001617766
Firm city
Westlake Village
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register