Skip to the content

FDA Medical Device Recalls (openFDA)

Karl Storz Endoscopy: FIVE S 3.5x65, Flexible Intubation Videoscope, REF: 091361-01 (1 Unit), and 091361-06 (6 Units)

Recall number
Z-0184-2022
Recalling firm
Karl Storz Endoscopy
Product
FIVE S 3.5x65, Flexible Intubation Videoscope, REF: 091361-01 (1 Unit), and 091361-06 (6 Units)
Status
Open, Classified
Initiated
2021-09-10
Posted
not on file
Terminated
not on file
Reason
During a regular requalification review, affected flexible intubation video endoscopes did not achieve the required sterility assurance level. There is a risk that the patient may be exposed to a higher risk of infection.
Root cause
Under Investigation by firm
Action
On 09/10/21 recall notices were mailed to customers. Action to be taken by the user 1. Immediately quarantine and discontinue use of the affected products. 2. Pass on this Urgent Field Safety Notice to all users of the products listed above and all other persons who need to be aware within your organization. 3. If you have distributed the affected products to third parties, please promptly forward this letter to the recipients of the products and indicate contact details of the recipient on the Acknowledgement and Response Form. 4. Return the Acknowledgment and Response Form by Fax or E-Mail to the contact listed below within 10 calendar days of receipt of this notice. 5. Get in touch with your KARL STORZ representative to return affected products and to discuss suitable alternative products. This notice must be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Complete and return the Customer Acknowledgement and Response Form.
Quantity
532
Distribution
US: SC, NH, UT, LA, FL, TX, MA, MO, VA, IA, CA, IL, NC, VT, AL, GA, PA, TN, WI, NY, CO, AZ, NV, ME, MS, DC, IN, KY, MD, OH, MI, WA, KS, OR, AR, MN, NJ, NM, WV, CT. OUS: Canada
Lot, serial or model codes
All LOTs with remaining shelf life
510(k) numbers
not on file
PMA numbers
not on file
Product code
CAL
Firm FEI
3007475226
Firm city
El Segundo
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register