Zimmer Biomet, Inc.: ROSA Surgical Device 2.5.8 ROSA Surgical Device is a computer-controlled electromechanical arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide. Guidance is based on a pre-operative plan developed with three-dimensional imaging software, and uses fiducial markers or optical registration. The system is intended for use by neurosurgeons to guide standard neurosurgical instruments. It is indicated for any neurosurgical condition in which the use of stereotactic surgery may be appropriate

Recall number
Z-0184-2018
Recalling firm
Zimmer Biomet, Inc.
Product
ROSA Surgical Device 2.5.8 ROSA Surgical Device is a computer-controlled electromechanical arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide. Guidance is based on a pre-operative plan developed with three-dimensional imaging software, and uses fiducial markers or optical registration. The system is intended for use by neurosurgeons to guide standard neurosurgical instruments. It is indicated for any neurosurgical condition in which the use of stereotactic surgery may be appropriate
Status
Terminated
Initiated
2014-09-05
Posted
not on file
Terminated
2020-06-09
Reason
The software issue described was corrected in the modification to the MXTTOUT controller parameter settings.
Root cause
Software design
Action
This is a retrospective report of a correction initiated on 05-Sept-2014. The software issue described was corrected in the modification to the MXTTOUT controller parameter settings. There were 8 affected devices at customer sites in the US. Field Service Technicians were deployed to the consignee locations to perform the system upgrade.
Quantity
29 (8 US and 21 OUS)
Distribution
Nationwide Distribution to AK, OH, TX, GA, and MI
Lot, serial or model codes
Serial #'s: RO10009, RO10011, RO10014, RO13023, RO13027, RO14031, RO14033 and RO14035
510(k) numbers
["K092239","K101791"]
PMA numbers
not on file
Product code
HAW
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register