FDA Medical Device Recalls (openFDA)
Philips Medical Systems: M1020B PulseOximetry Module Philips FAST Sp02 and Pulse Oximetry Nellcor OxiMax Compatible used with the IntelliVue or M3/M4 patient monitor M1002
- Recall number
- Z-0184-2007
- Recalling firm
- Philips Medical Systems
- Product
- M1020B PulseOximetry Module Philips FAST Sp02 and Pulse Oximetry Nellcor OxiMax Compatible used with the IntelliVue or M3/M4 patient monitor M1002
- Status
- Terminated
- Initiated
- 2006-09-21
- Posted
- 2006-11-16
- Terminated
- 2020-08-11
- Reason
- Unexpected pulse oximetry (Sp02) readings (100%) over time when sensor is not attached to patient. In addition, with a sensor attached to a patient, an incorrect high Sp02 may be displayed when a pulse rate is in a range of 185BPM
- Root cause
- Other
- Action
- Philips issued a Customer Letter and Urgent Device Correction letter to consignees on 09/21/06. Users were notified of the problem and provided with Instructions to Mitigate Risk. A software upgrade to: version D.00.61 and Sp02 FW A.01.46 is mandatory.
- Quantity
- 442 units
- Distribution
- Worldwide, including USA, Canada, Australia, China, Hong Kong, India, Indonesia, Japan, Phillipines, Singapore, Taiwan, Thailand, Vietnam, Argentina, Brazil, Chile, Colombia, El Salvador, Jamaica, Mexico, Nicaragua. Panama, Paraguay, Peru, Venezuela, Austria, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hungary, Germany, Israel, Italy, Jordan, Kenya, Kuwait, Lebanon, Lithuania, Luxembourg, Macedonia, Mauritania, Mauritius, Netherlands, Norway, Oman, Slovenia, South Africa, Spain, Switzerland, Turkey, UAE, UK, Yemen, Zambia, Zimbabwe
- Lot, serial or model codes
- M1020B #AO01 AND #A02 with S/N DE524xxxxx M102B #A01 and #A02 with S/N DE612xxxxx With Sp02 Firmware (Fw) Rev: Sp02 DSP Fw-Rev. A.01.42
- 510(k) numbers
- ["K033715"]
- PMA numbers
- not on file
- Product code
- DQA
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28