Skip to the content

FDA Medical Device Recalls (openFDA)

Focus Diagnostics Inc: West Nile Virus IgG DxSelect" kit Catalog No. EL0300G Kit Lots 122150, 122155, 122641, 122650 and 122859 that includes West Nile Virus IgG Antigen Wells, Catalog Number EL0351, Lot Number 121117. 510(k) No. K031953 The Focus Diagnostics West Nile Virus ELISA IgG is intended for qualitatively detecting IgG antibodies to West Nile virus in human serum. In conjunction with the Focus Diagnostics West Nile Virus IgM Capture ELISA, the test is indicated for testing persons having symptoms of meningoencephalitis, as an aid in the presumptive laboratory diagnosis of West Nile virus infection.

Recall number
Z-0182-2013
Recalling firm
Focus Diagnostics Inc
Product
West Nile Virus IgG DxSelect" kit Catalog No. EL0300G Kit Lots 122150, 122155, 122641, 122650 and 122859 that includes West Nile Virus IgG Antigen Wells, Catalog Number EL0351, Lot Number 121117. 510(k) No. K031953 The Focus Diagnostics West Nile Virus ELISA IgG is intended for qualitatively detecting IgG antibodies to West Nile virus in human serum. In conjunction with the Focus Diagnostics West Nile Virus IgM Capture ELISA, the test is indicated for testing persons having symptoms of meningoencephalitis, as an aid in the presumptive laboratory diagnosis of West Nile virus infection.
Status
Terminated
Initiated
2012-10-03
Posted
2012-10-31
Terminated
2012-12-04
Reason
The firm recalled due to higher reactivity with some samples in the effected lots, which may reflect a higher false positive rate.
Root cause
Nonconforming Material/Component
Action
Focus Diagnostics Inc., sent a Urgent Medical Device Recall letter dated October 3, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Our records indicate that you received at least one of the effected lots in the month of August 2012. Please document the number of kits in your inventory on the table below, even if zero (0), and fax to 562-240-6526. For further questions please call (714) 220-1900.
Quantity
425 units
Distribution
Worldwide Distribution--US (nationwide) and the country of: Canada, Deuche, France, Greece, Mexico, Israel, and Japan.
Lot, serial or model codes
Catalog No. EL0300G Kit Lots 122150, 122155, 122641, 122650 and 122859 that includes West Nile Virus IgG Antigen Wells, Catalog Number EL0351, Lot Number 121117.
510(k) numbers
["K031953"]
PMA numbers
not on file
Product code
NOP
Firm FEI
2023365
Firm city
Cypress
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register