FDA Medical Device Recalls (openFDA)
Stryker Communications Corp: Circlip component of the Stryker Flat Panel and Navigation Arm System. Stryker Flat Panel and Navigation Arm System are intended for use as a ceiling-mounted device to support a medical grade flat panel monitor or camera in the patient area. The Circlip within the flat panel and navigation arms is part of the mechanical stop designed to prevent the arm from rotating infinitely. It also prevents the cables routed through the arm from becoming twisted during routine movement and rotation of the arm.
- Recall number
- Z-0182-2009
- Recalling firm
- Stryker Communications Corp
- Product
- Circlip component of the Stryker Flat Panel and Navigation Arm System. Stryker Flat Panel and Navigation Arm System are intended for use as a ceiling-mounted device to support a medical grade flat panel monitor or camera in the patient area. The Circlip within the flat panel and navigation arms is part of the mechanical stop designed to prevent the arm from rotating infinitely. It also prevents the cables routed through the arm from becoming twisted during routine movement and rotation of the arm.
- Status
- Terminated
- Initiated
- 2008-09-05
- Posted
- 2008-10-22
- Terminated
- 2009-10-02
- Reason
- Circlip component used to suspend flat panel and navigation arm system may become dislodged.
- Root cause
- Device Design
- Action
- Notification letter was sent to consignees September 5, 2008. Field service will be performed at all affected sites. Old Circlips will be replaced with new clamps that do not allow the arm to fall. Contact Stryker Communications Corp. at 1-972-410-7310 for assistance.
- Quantity
- 705 units
- Distribution
- Nationwide and International Distribution
- Lot, serial or model codes
- Part numbers: 0682-000-005 (A-O), 0682-000-200, 0682-000-270, 0682-000-280, 0682-000-281, 0682-000-850, and 0682-000-851. All units distributed between December 23, 2003 and September 30, 2006.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FQO
- Firm FEI
- 3002975342
- Firm city
- Flower Mound
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as