Skip to the content

FDA Medical Device Recalls (openFDA)

Olsen Medical: MIDAS TOUCH Electrosurgical/102 mm (4 inch) Blade Electrodes, Disposable, Sterile Single Use Device, Part #30-0012 and #30-0002. The electrodes are individually packaged in Tyvek bags with 12 packages per case.

Recall number
Z-0182-2007
Recalling firm
Olsen Medical
Product
MIDAS TOUCH Electrosurgical/102 mm (4 inch) Blade Electrodes, Disposable, Sterile Single Use Device, Part #30-0012 and #30-0002. The electrodes are individually packaged in Tyvek bags with 12 packages per case.
Status
Terminated
Initiated
2006-08-29
Posted
2006-11-15
Terminated
2011-07-26
Reason
Testing conducted by the firm indicates that electrodes identified with certain part numbers may fall out from the electrosurgical pencil or handle presenting the potential for harm or injury to the patient and/or user.
Root cause
Other
Action
The firm issued an ''Urgent: Device Recall'' notification to their customers via letter and E-mail on 8/29/2006 and 8/30/2006. The letter informs the customers of this problem and asks that the customers contact their sub-accounts and notify them of the problem if they have further distributed any of the suspect part numbers. The letter also asks that the distributor/customer(s) have the subaccounts discontinue use of the product; and return the held product to the distributor (who in turn, should return the product to Olsen Medical for reimbursement). The customers are asked to complete and fax back to the recalling firm, an attached Acknowledgement form documenting receipt of the recall notification by the customer/employee with name and date received. The firm also asked that the customer supply information regarding the amount of inventory /shipped on a checklist. The amounts received are broken down on the Checklist by Lot Number with the amount of product in inventory and the amount of product shipped. These amounts are reported by the responders by ''pieces and cases'' on hand at the time that the notice was received.
Quantity
81 cases/972 pieces
Distribution
Worldwide, including USA, Japan, and New Zealand.
Lot, serial or model codes
4 inch Standard Electrodes - Model Number: 30-0002, Lot Numbers: 110201, 116228, 117331, 117985, 120959 and 000501; and 4 inch Modified Electrodes - Part Number: 30-0012, Lot Numbers: 110101, 113523, 113877, 114428, 114527, 115294, 115420, 119874, 120223, 120732 and 120883.
510(k) numbers
["K973234"]
PMA numbers
not on file
Product code
JOS
Firm FEI
3000719969
Firm city
Louisville
Firm state
KY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register