Skip to the content

FDA Medical Device Recalls (openFDA)

OriGen Biomedical, Inc.: Reinforced Dual Lumen Cannula, 16Fr., REF: VV16F, Sterile EO Indications: The OriGen Reinforced Dual Lumen Cannula is indicated for use as a single cannula for both venous drainage and re-infusion of blood in the internal jugular vein during extracorporeal life support procedures.

Recall number
Z-0180-2019
Recalling firm
OriGen Biomedical, Inc.
Product
Reinforced Dual Lumen Cannula, 16Fr., REF: VV16F, Sterile EO Indications: The OriGen Reinforced Dual Lumen Cannula is indicated for use as a single cannula for both venous drainage and re-infusion of blood in the internal jugular vein during extracorporeal life support procedures.
Status
Terminated
Initiated
2018-08-21
Posted
not on file
Terminated
2021-07-15
Reason
These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip
Root cause
Component design/selection
Action
On 08/21/18, the firm, OriGen, emailed a recall notice dated 8/17/2018 to customers informing them of a nationwide recall of RDLC ECMO catheters. Customers were advised to return affected product to the firm. Customers with additional questions were advised to call 1-800-233-9014 or 1-512-474-7278. The firm also sent customers Technical Bulletin (TB) 1411 - Revision Feb 2018, Subject: Radiopacity of the Reinforced Dual Lumen Catheter (RDLC), Date: July 2015, and an Acknowledgment Receipt form to complete and return. On 09/12/18, the firm, Origen, sent another recall notice, dated 9/12/2018 which included additional affected lot numbers. TB1411 was sent to customers three additional times (Version/Customer Notification Date): R00/11/10/14, R01/07/22/15, and Revision Feb 2018/02/12/18. Consumers with questions may contact the company via telephone at 1-800 233 9014 between the hours of 08:00 and 16:30 (CDT). Customers outside the USA may contact OriGen at +1.512.474.7278. Customers may also contact the company via e-mail at ecmo@origen.com.
Quantity
452
Distribution
Worldwide Distribution: U.S.(nationwide) to states of: IN, NY, CA, MA, TX, OH, KY, MI, MS, NM, NC, VA, WI, TN, CT, FL, AR, MO, GA, LA, DC, OR, MN, SC, OK, AZ, UT, RI, SD, WA, PA, AL, MD, IL, HI; and countries (OUS) of: Argentina, Austria, Belgium, CANADA, Colombia, EU, Germany, India, Italy, Japan, Kingdom of Saudi Arabia, Netherlands, New Zealand, Poland, Qatar, Romania, South Africa, SWEDEN, Thailand, UK, United Arab Emirates, and VIETNAM.
Lot, serial or model codes
UDI 00816203022073, Lot: N18404, N18446, N18640, N18640-1, R19208
510(k) numbers
["K113869"]
PMA numbers
not on file
Product code
DWF
Firm FEI
1000160256
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register