Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific CRM Corp: Guidant LATITUDE Communicator, model 6481, Boston Scientific 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. The LATITUDE Patient Management system enables physicians to periodically monitor both patient and device status remotely. The LATITUDE Patient Management system is intended for use to remotely communicate with a compatible Guidant or Boston Scientific pulse generator and transfer data to a central database.

Recall number
Z-0180-2010
Recalling firm
Boston Scientific CRM Corp
Product
Guidant LATITUDE Communicator, model 6481, Boston Scientific 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. The LATITUDE Patient Management system enables physicians to periodically monitor both patient and device status remotely. The LATITUDE Patient Management system is intended for use to remotely communicate with a compatible Guidant or Boston Scientific pulse generator and transfer data to a central database.
Status
Terminated
Initiated
2009-08-07
Posted
2009-11-12
Terminated
2012-12-04
Reason
The L ATITUDE communicator (Model 6481) for a patient may not have been able to receive measurements from their weight scale or blood pressure monitor afer initial setup. The issue has been resolved. If any patient weight or blood pressure readings are stored in the patient's sensor, they will be uploaded to LATITUDE after the patient's communicator calls in to Boston Scientific's servers. It
Root cause
Software design
Action
Boston Scientific used phone calls as the delivery method for these communications. The communication began August 6, 2009 for the original 6 patients affected. Notifications via phone calls to additional patients affected began on August 17, 2009. Direct questions about the recall to the LATITUDE Customer Service by calling 1-651-582-4000.
Quantity
97
Distribution
Nationwide Distribution -- AZ, AR, CA, CT, FL, GA, IL, IN, MA, MO, NV, NJ, NY, OH, OR, PA, SC, TX, UT, VA and WA.
Lot, serial or model codes
SERIAL NUMBERS:  029251, 064452, 059718, 064001, 056441, 059851, 070188, 072291, 312706, 305339, 312762, 004610, 024762, 049668, 061205, 067781, 051788, 056069, 068447, 069808, 008653, 068085, 057090, 023576, 028974, 017638, 603254, 064457, 069473, 069902, 071065, 072021, 073621, 073837, 312707, 605519,  603234, 603788, 072396, 059662, 056346, 301645, 312781, 307327, 067067,  055853, 060274, 605304, 305344, 302301, 605981, 009271, 056500, 300895,  602642, 030830, 301117, 055001, 049660, 047455, 056923, 309556, 044034,  063667, 005931, 067765, 063012, 050130, 051267, 055558, 060325, 060973,  063034, 306808, 306852, 309355, 041156, 062922, 063064, 061177, 300499,  039398, 039401, 039403, 060619, 065476, 058913, 056802, 073812  024833, 073560, 606104, 606288, 606604, 606826, 607180 608231 and 607180.
510(k) numbers
not on file
PMA numbers
["P910077S092"]
Product code
LWS
Firm FEI
2124215
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register