Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: Allen Life Assist Beach Chair, BCSSP-BEACH CHAIR , REF A-91501, #104002. For use with orthopaedic procedures using the Fowlers or beach chair position.

Recall number
Z-0179-2023
Recalling firm
Baxter Healthcare Corporation
Product
Allen Life Assist Beach Chair, BCSSP-BEACH CHAIR , REF A-91501, #104002. For use with orthopaedic procedures using the Fowlers or beach chair position.
Status
Open, Classified
Initiated
2022-09-20
Posted
2022-11-03
Terminated
not on file
Reason
the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.
Root cause
Other
Action
The firm issued an Urgent Medical Device Correction notice to their consignee on 09/20/2022 by email. The notice explained the problem, risk, and requested the following: "Actions to be taken by Customer: 1. Identify the impacted lot number listed on page 1 and stop any use of the product immediately. 2. Hillrom will contact you to perform a correction of your impacted device as soon as possible." The firm issued an Urgent Medical Device Recall notice to their consignees on 09/27/2022 by FedEx 2nd day delivery. The notice explained the problem, risk, and requested the following: "Actions to be taken by Customer: 1. Identify the impacted lot numbers listed in the above table and replace with the enclosed device that accompanies this notification. 2. Please return the impacted unit per RMA instructions attached within 10 business days. Actions to be taken by Distributor: 1. Please forward this notification to your end users along with replacement parts accompanying this notification. 2. Please request return of the impacted devices from your end users and return the devices to Hillrom, per RMA instructions attached." Both notices provided the following contact information: "Contact reference person If you have any questions regarding this Urgent Medical Device Correction, please contact Hillrom Technical Support at 800-445-3720 or technical.support@hillrom.com"
Quantity
1 unit
Distribution
MI, BA, WI, WA, TX, and Germany, Japan, South Korea, Sweden, South Africa
Lot, serial or model codes
UDI/DI 00615521004125, Lot Number: 22-123
510(k) numbers
not on file
PMA numbers
not on file
Product code
GDC
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register