Leica Microsystems, Inc.: Thyroglobulin (1D4) is a mouse anti-human monoclonal antibody produced as a tissue culture supernatant, and supplied in Tris buffered saline with carrier protein, containing 0.35% ProClin 950 as a preservative. Total volume = 7 mL. This reagent is for in vitro diagnostic use. Thyroglobulin (1D4) monoclonal antibody is intended to be used for the qualitative identification by light microscopy of thyroglobulin in formalin-fixed, paraffin-embedded tissue by immunohistochemical staining using an automated Bond system. The clinical interpretation of any staining or its absence should be complemented by morphological studies and proper controls and should be evaluated within the context of the patient's clinical history and other diagnostic tests by a qualified pathologist.

Recall number
Z-0179-2015
Recalling firm
Leica Microsystems, Inc.
Product
Thyroglobulin (1D4) is a mouse anti-human monoclonal antibody produced as a tissue culture supernatant, and supplied in Tris buffered saline with carrier protein, containing 0.35% ProClin 950 as a preservative. Total volume = 7 mL. This reagent is for in vitro diagnostic use. Thyroglobulin (1D4) monoclonal antibody is intended to be used for the qualitative identification by light microscopy of thyroglobulin in formalin-fixed, paraffin-embedded tissue by immunohistochemical staining using an automated Bond system. The clinical interpretation of any staining or its absence should be complemented by morphological studies and proper controls and should be evaluated within the context of the patient's clinical history and other diagnostic tests by a qualified pathologist.
Status
Terminated
Initiated
2014-04-18
Posted
2014-11-06
Terminated
2017-08-14
Reason
Bond Ready to Use Primary Antibody Thyroglobulin does not function as intended up to the expiry date on the product labelling. This was detected through an on-going stability program. There is a link between the age of the product and staining intensity.
Root cause
Unknown/Undetermined by firm
Action
Leica Biosystems sent a Medical Device Recall Notification letter dated April 18, 2014, to all affected customers. The letter included instructions for customers to: 1) assess any results previously obtained by using the recalled products; 2) identify and destroy any unused or partly used recalled products or confirm that the recalled products are no longer in stock; and, 3) complete and return the attached Medical Device Recall Notification Acknowledgement Form. Customers with questions were instructed to contact Leica Microsystems by e-mail at LMGRA@leica-microsystems.com. For questions regarding this recall call 847-405-5413.
Quantity
180 units
Distribution
Worldwide Distribution - USA including AR, AZ, CA, CO, CT, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, MT, MY, NC, NH, NJ, NY, OH, ON, OR, PA, TN, TX, UT, VA, WA, WI and the District of Columbia, Internationally to Belgium, Bulgaria, France, Italy, Pakistan, Philippines, Russia, Saudi Arabia, Spain, Turkey, and the United Kingdom.
Lot, serial or model codes
Product Code: PA0025; Lot Numbers and Expiration Dates: 24963, Expires 17 DEC 2014; 24228, Expires 15 OCT 2014; 23882, Expires 10 SEP 2014; 23366, Expires 23 JUL 2014
510(k) numbers
not on file
PMA numbers
not on file
Product code
NJT
Firm FEI
3000210133
Firm city
Buffalo Grove
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register