FHC, Inc.: FHC microTargeting Power Assist System used with the micro Targeting Drive System, models: 66-DA-ME; 66EL-MC; 66EL-MS;66EL-MS-02; 66-EL-MS-03; 66-EL-RM; MT-DS-AM; MT-DS-MD.

Recall number
Z-0179-06
Recalling firm
FHC, Inc.
Product
FHC microTargeting Power Assist System used with the micro Targeting Drive System, models: 66-DA-ME; 66EL-MC; 66EL-MS;66EL-MS-02; 66-EL-MS-03; 66-EL-RM; MT-DS-AM; MT-DS-MD.
Status
Terminated
Initiated
2005-10-31
Posted
2005-11-23
Terminated
2006-06-07
Reason
Power Assist microTargeting System may advance beyond intended target
Root cause
Other
Action
FHC Inc. and Medtronic notified customers by letter and instructs the customer to return their Motor Option for upgrade to include the following components: 1) Handheld Remote Control, 2) Motor Controller Unit with Display, 3) Motor, and 4) Medical Power Adapter. (See pictures at the end of this communication). Until the Motor Option has been updated, instructions to customers are: Prior to each clinical use: Ensure the microTargeting Drive system is calibrated before each clinical use per the User Guide. Check the microTargeting Drive system functionality over a range of at least 10mm movement. Should any performance indicating either a stall or unintended movement be observed, immediately unplug the 12V Medical Power Adapter, remove the MicroTargeting Drive Motor Option from the system, and continue using the microTargeting Drive in the manual mode. During clinical use: Should any performance indicating either a stall or unintended movement be observed, immediately unplug the 12V Medical Power Adapter, remove the MicroTargeting Drive Motor Option from the system, and continue using the microTargeting Drive in the manual mode. Every 100 uses or annually: Have your microTargeting Drive system checked by the manufacturer per the User Guide. Have your equipment support department check or tighten the MicroTargeting Drive Motor Option Remote Control Knob set screws, tighten as needed, and then re-calibrate.
Quantity
70 units
Distribution
Nationwide Foreign: South Africa, Taiwan, Hong Kong, China, France
Lot, serial or model codes
N/A
510(k) numbers
["K011992"]
PMA numbers
not on file
Product code
HAW
Firm FEI
3002250546
Firm city
Bowdoinham
Firm state
ME
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register