FHC, Inc.: FHC microTargeting Power Assist System used with the micro Targeting Drive System, models: 66-DA-ME; 66EL-MC; 66EL-MS;66EL-MS-02; 66-EL-MS-03; 66-EL-RM; MT-DS-AM; MT-DS-MD.
- Recall number
- Z-0179-06
- Recalling firm
- FHC, Inc.
- Product
- FHC microTargeting Power Assist System used with the micro Targeting Drive System, models: 66-DA-ME; 66EL-MC; 66EL-MS;66EL-MS-02; 66-EL-MS-03; 66-EL-RM; MT-DS-AM; MT-DS-MD.
- Status
- Terminated
- Initiated
- 2005-10-31
- Posted
- 2005-11-23
- Terminated
- 2006-06-07
- Reason
- Power Assist microTargeting System may advance beyond intended target
- Root cause
- Other
- Action
- FHC Inc. and Medtronic notified customers by letter and instructs the customer to return their Motor Option for upgrade to include the following components: 1) Handheld Remote Control, 2) Motor Controller Unit with Display, 3) Motor, and 4) Medical Power Adapter. (See pictures at the end of this communication). Until the Motor Option has been updated, instructions to customers are: Prior to each clinical use: Ensure the microTargeting Drive system is calibrated before each clinical use per the User Guide. Check the microTargeting Drive system functionality over a range of at least 10mm movement. Should any performance indicating either a stall or unintended movement be observed, immediately unplug the 12V Medical Power Adapter, remove the MicroTargeting Drive Motor Option from the system, and continue using the microTargeting Drive in the manual mode. During clinical use: Should any performance indicating either a stall or unintended movement be observed, immediately unplug the 12V Medical Power Adapter, remove the MicroTargeting Drive Motor Option from the system, and continue using the microTargeting Drive in the manual mode. Every 100 uses or annually: Have your microTargeting Drive system checked by the manufacturer per the User Guide. Have your equipment support department check or tighten the MicroTargeting Drive Motor Option Remote Control Knob set screws, tighten as needed, and then re-calibrate.
- Quantity
- 70 units
- Distribution
- Nationwide Foreign: South Africa, Taiwan, Hong Kong, China, France
- Lot, serial or model codes
- N/A
- 510(k) numbers
- ["K011992"]
- PMA numbers
- not on file
- Product code
- HAW
- Firm FEI
- 3002250546
- Firm city
- Bowdoinham
- Firm state
- ME
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28