Zimmer Biomet, Inc.: Rosa Spine 1.0.2 Stereotaxic Instrument Computer-Assisted Surgical Device The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by surgeons to guide standard neurosurgical instruments during spine surgery. Guidance is based on an intra-operative plan developed with three-dimensional imaging software provided that the required fiducial markers and rigid patient anatomy can be identified on 3D CT scans. The device is indicated for the placement of pedicle screws in lumber vertebrae with a posterior approach.
- Recall number
- Z-0177-2018
- Recalling firm
- Zimmer Biomet, Inc.
- Product
- Rosa Spine 1.0.2 Stereotaxic Instrument Computer-Assisted Surgical Device The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by surgeons to guide standard neurosurgical instruments during spine surgery. Guidance is based on an intra-operative plan developed with three-dimensional imaging software provided that the required fiducial markers and rigid patient anatomy can be identified on 3D CT scans. The device is indicated for the placement of pedicle screws in lumber vertebrae with a posterior approach.
- Status
- Terminated
- Initiated
- 2015-10-29
- Posted
- not on file
- Terminated
- 2020-06-16
- Reason
- A design change was initiated to update ROSA Spine 1.0.2 to version ROSA Spine 1.0.2.16 to resolve software bugs and improve usability and stability of the ROSA Spine device.
- Root cause
- Software design
- Action
- Zimmer Biomet issued a retrospective report of a correction taken on 29-Oct-2015. The software issue described was corrected in the new software version 1.0.2.16. There were 2 copies of the affected software in distribution. Field Service Technicians were deployed to the consignee locations to perform the system upgrade. 100% of affected devices were updated with the software solution. No further action is required.
- Quantity
- 2
- Distribution
- Worldwide Distribution including France and Germany
- Lot, serial or model codes
- Serial #'s; SP14003 and SP14005
- 510(k) numbers
- ["K151511"]
- PMA numbers
- not on file
- Product code
- OLO
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28