Datascope Corp: Passport 2 with View 12 ECG Analysis Module. It is a transportable, multi-parameter physiological monitor designed to monitor and display the following physiological parameters: ECG, Heart Rate derived from selected sources (Sp)2, ECG, IBP and NIBP), SpO2 level, ST Segment, Arrhythmia, Blood pressure (both invasive and non-invasive), Respiration rate (dirived from ECG or CO2), inspired or expired CO2, temperature , and gases (i.e. five anesthetsia gases, O2, N2O, and CO2)

Recall number
Z-0177-06
Recalling firm
Datascope Corp
Product
Passport 2 with View 12 ECG Analysis Module. It is a transportable, multi-parameter physiological monitor designed to monitor and display the following physiological parameters: ECG, Heart Rate derived from selected sources (Sp)2, ECG, IBP and NIBP), SpO2 level, ST Segment, Arrhythmia, Blood pressure (both invasive and non-invasive), Respiration rate (dirived from ECG or CO2), inspired or expired CO2, temperature , and gases (i.e. five anesthetsia gases, O2, N2O, and CO2)
Status
Terminated
Initiated
2005-09-07
Posted
2005-11-23
Terminated
2007-06-06
Reason
Software anomalies which include shut down while printing, invalid diastolic blood pressure display, and incomplete and inaccurate interpretation reports of 12-Lead ECG data.
Root cause
Other
Action
Datascope sent out a Field Correction notice on 9/7/2005 to all Hospital administrators who have the monitors. This was sent out via certified mail, return receipt requested. A Datascope Service Rep will conduct and document the corrective action during a visit to the customer.
Quantity
979 monitors
Distribution
The monitors have been distributed domestically and internationally to hospitals and clinics. There are 10 Passport 2 monitors distributed to Naval Hospital, Camp Pendleton, CA. There are 12 Spectrum monitors distributed to two US Govt facilities: VA Medical Center, Hines, IL and VA Medical Center, Syracuse, NY.
Lot, serial or model codes
Passport 2 Monitors with software versions: S.10, T.15, T.31, V.07, W.09, W.13, W.14.
510(k) numbers
["K020550"]
PMA numbers
not on file
Product code
DRT
Firm FEI
2221819
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register