FDA Medical Device Recalls (openFDA)
Datascope Corp: Passport 2 with View 12 ECG Analysis Module. It is a transportable, multi-parameter physiological monitor designed to monitor and display the following physiological parameters: ECG, Heart Rate derived from selected sources (Sp)2, ECG, IBP and NIBP), SpO2 level, ST Segment, Arrhythmia, Blood pressure (both invasive and non-invasive), Respiration rate (dirived from ECG or CO2), inspired or expired CO2, temperature , and gases (i.e. five anesthetsia gases, O2, N2O, and CO2)
- Recall number
- Z-0177-06
- Recalling firm
- Datascope Corp
- Product
- Passport 2 with View 12 ECG Analysis Module. It is a transportable, multi-parameter physiological monitor designed to monitor and display the following physiological parameters: ECG, Heart Rate derived from selected sources (Sp)2, ECG, IBP and NIBP), SpO2 level, ST Segment, Arrhythmia, Blood pressure (both invasive and non-invasive), Respiration rate (dirived from ECG or CO2), inspired or expired CO2, temperature , and gases (i.e. five anesthetsia gases, O2, N2O, and CO2)
- Status
- Terminated
- Initiated
- 2005-09-07
- Posted
- 2005-11-23
- Terminated
- 2007-06-06
- Reason
- Software anomalies which include shut down while printing, invalid diastolic blood pressure display, and incomplete and inaccurate interpretation reports of 12-Lead ECG data.
- Root cause
- Other
- Action
- Datascope sent out a Field Correction notice on 9/7/2005 to all Hospital administrators who have the monitors. This was sent out via certified mail, return receipt requested. A Datascope Service Rep will conduct and document the corrective action during a visit to the customer.
- Quantity
- 979 monitors
- Distribution
- The monitors have been distributed domestically and internationally to hospitals and clinics. There are 10 Passport 2 monitors distributed to Naval Hospital, Camp Pendleton, CA. There are 12 Spectrum monitors distributed to two US Govt facilities: VA Medical Center, Hines, IL and VA Medical Center, Syracuse, NY.
- Lot, serial or model codes
- Passport 2 Monitors with software versions: S.10, T.15, T.31, V.07, W.09, W.13, W.14.
- 510(k) numbers
- ["K020550"]
- PMA numbers
- not on file
- Product code
- DRT
- Firm FEI
- 2221819
- Firm city
- Mahwah
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as