Medical Industries America Inc: Aeroneb Go Micropump Nebulizer Featuring OnQ electronic micropump, RX, Manufactured by Medical Industries America Inc., Adel, IA. The product is sold individually or in master packs of 8 under the following model numbers (all using the same box label): Model 7000 - Consisting of the Aeroneb Go, battery controller, battery controller cable, 3 AA alkaline batteries, mask elbow, mouthpiece, carry bag, and user manual; Model 7030 - Consisting of the Aeroneb Go, AC wall adapter, mask elbow, mouthpiece, carry bag, and user manual; Model 7070 - Consisting of the Aeroneb Go, battery controller, battery controller cable, 3 AA alkaline batteries, AC wall adapter, mask elbow, mouthpiece, carry bag, and user manual; Model 7000-1 - Consisting of the Aeroneb Go, battery controller, battery controller cable, 3 AA alkaline batteries, European mask elbow, mouthpiece, carry bag, and user manual; Model 7030-1 - Consisting of the Aeroneb Go, AC wall adapter, European mask elbow, mouthpiece, carry bag, and user manual; Model 7070-1 - Consisting of the Aeroneb Go, battery controller, battery controller cable, 3 AA alkaline batteries, AC wall adapter, European mask elbow, mouthpiece, carry bag, and user manual. 7000LINCARE - Consisting of the Aeroneb Go, battery controller with a Lincare private label cover plate, battery controller cable, 3 AA alkaline batteries, mask elbow, mouthpiece, carry bag, and user manual; 7070LINCARE - Consisting of the Aeroneb Go, battery controller with the Lincare private label cover plate, battery controller cable, 3 AA alkaline batteries, AC wall adapter, mask elbow, mouthpiece, carry bag, and user manual.

Recall number
Z-0177-05
Recalling firm
Medical Industries America Inc
Product
Aeroneb Go Micropump Nebulizer Featuring OnQ electronic micropump, RX, Manufactured by Medical Industries America Inc., Adel, IA. The product is sold individually or in master packs of 8 under the following model numbers (all using the same box label): Model 7000 - Consisting of the Aeroneb Go, battery controller, battery controller cable, 3 AA alkaline batteries, mask elbow, mouthpiece, carry bag, and user manual; Model 7030 - Consisting of the Aeroneb Go, AC wall adapter, mask elbow, mouthpiece, carry bag, and user manual; Model 7070 - Consisting of the Aeroneb Go, battery controller, battery controller cable, 3 AA alkaline batteries, AC wall adapter, mask elbow, mouthpiece, carry bag, and user manual; Model 7000-1 - Consisting of the Aeroneb Go, battery controller, battery controller cable, 3 AA alkaline batteries, European mask elbow, mouthpiece, carry bag, and user manual; Model 7030-1 - Consisting of the Aeroneb Go, AC wall adapter, European mask elbow, mouthpiece, carry bag, and user manual; Model 7070-1 - Consisting of the Aeroneb Go, battery controller, battery controller cable, 3 AA alkaline batteries, AC wall adapter, European mask elbow, mouthpiece, carry bag, and user manual. 7000LINCARE - Consisting of the Aeroneb Go, battery controller with a Lincare private label cover plate, battery controller cable, 3 AA alkaline batteries, mask elbow, mouthpiece, carry bag, and user manual; 7070LINCARE - Consisting of the Aeroneb Go, battery controller with the Lincare private label cover plate, battery controller cable, 3 AA alkaline batteries, AC wall adapter, mask elbow, mouthpiece, carry bag, and user manual.
Status
Terminated
Initiated
2004-09-07
Posted
2004-11-05
Terminated
2005-08-24
Reason
The nebulizer can either fail or provide a low flow (partial dose) of medication
Root cause
Other
Action
The recalling firm initiated their action via telephone on 9/7/04 to the care providers to inform them of the product problem and that they would soon be receiving additional information about the recall. Between 9/24-28/04, the firm issued letters to their care providers with specific instructions as to how to decide what individuals would or would not need a backup generator in their possession. They were also instructed to pass the information onto the end user. On 10/1/04, the firm called the care providers and subsequently issued second letters or postcards to them to correct the toll free number that had been misprinted on the literature and letter.
Quantity
17,464 units
Distribution
Distribution was made to home health care providers and pharmacies nationwide. Foreign distribution was made to Italy, Greece, France, Poland, Japan, Hong Kong, Argentina, Australia, Israel, and Singapore.
Lot, serial or model codes
040120, 040127, 040129, 040130, 040201, 040202, 040203, 040205, 040209, 040210, 040216, 040219, 040227, 040301, 040303, 040305, 040311, 040317, 040406, 040407, 040415, 040427, 040421, 040428, 040506, 040721, 040723, 040726, 040730, 080504, 080904, 081104, 081604, 081704, 081904, 082004, 082304, 082504, 082704, 083004, 083104, 080304
510(k) numbers
["K032849"]
PMA numbers
not on file
Product code
CAF
Firm FEI
1931654
Firm city
Adel
Firm state
IA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register