Bien-Air Surgery Sa: Bien-Air Surgery REF 1600407-001 SN **** FOOTCTRL OSSEODOC - Product Usage: intended to convert electrical energy to mechanically drive handpieces and microsaws used to cut and shape bones.

Recall number
Z-0176-2021
Recalling firm
Bien-Air Surgery Sa
Product
Bien-Air Surgery REF 1600407-001 SN **** FOOTCTRL OSSEODOC - Product Usage: intended to convert electrical energy to mechanically drive handpieces and microsaws used to cut and shape bones.
Status
Open, Classified
Initiated
2020-09-11
Posted
not on file
Terminated
not on file
Reason
The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor.
Root cause
Process control
Action
On 09/14/2020, the firm send out an "URGENT: Medical Device Correction" notification informing customer of the potential for the magnet located inside the foot pedal displaces or comes off during a surgery and therefore may compromise the device's safety. The use of the incriminated devices may entail a risk of unattended start or unstoppable motor when acting on the foot pedal. The Recalling Firm is informing customers of a curative and a corrective action to prevent the magnet from coming off: -Curative Action: 1) Inspect your inventory for the product numbers/serial numbers above. 2) Quarantine any of the affected products until performing the following actions 3) Open the foot pedal rocker to have access to the incriminated magnet by pulling the black part 4) Carefully inspect the position of the magnet using illustrations provided in the Customer Notification If the magnet is partially or full off, Customers are instructed to keep the foot pedals in quarantine (Do Not Use It Anymore), and contact the Recalling Firm's repair center and send the foot pedal back for repair. 5) If the magnet is well position, then carefully clean the plastic area around the magnet with a disinfectant wipe (e.g. alcoholic solution) and a swab to correctly clean the corners 6) Make sure the surface is clean and dry before applying an adhesive tape (e.g. Scotch, width between 10-15mm, length between 55-60mm) as described in the illustrations provide in the Customer Notification THE FOLLOWING ACTIONS MUST BE PERFORMED BEFORE EACH SURGERY 7) Carefully inspect that the adhesive tape is correctly holding the magnet. If needed, remove the tape and restart the process from step 4. 8) Close the foot pedal rocker, connect the foot pedal cable to the control unit and switch it ON 9) Proceed with the following tests before performing the surgery a. Select one motor (BASCH, 80K, NANO, RAPIDO) or one handpiece (S120, OSSEOSTAP, PERFO) depending on your needs b. Hold the motor
Quantity
Total US = 1120 foot pedals
Distribution
US Nationwide distribution including in the states of CA, MI and MN.
Lot, serial or model codes
Model 1600407 / Catalog number 1600407-001 Serial Numbers: 16C0017 16F0013 16I0009 16K0017 17F0039 17I0015 17L0013
510(k) numbers
["K792159"]
PMA numbers
not on file
Product code
ERL
Firm FEI
3003800209
Firm city
Le Noirmont
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register