Datascope Corp: Spectrum Patient Monitor, Physiological, Patient with arrhythmia detection or alamrs. Monitor can monitor, display, trend and print a patient''s physiological parameters. The device has a 12.1 inch color display and has a standard configuration of a 3 or 5 lead ECG, Masimo SET SpO2, non-invasive blood pressure, respiration, continuous temperature and IV Drug Calculations. Optional digital displays are provided for invasive blood pressure (up to four) anesthetic agents, O2 and NO2, ST, and CO2.

Recall number
Z-0176-06
Recalling firm
Datascope Corp
Product
Spectrum Patient Monitor, Physiological, Patient with arrhythmia detection or alamrs. Monitor can monitor, display, trend and print a patient''s physiological parameters. The device has a 12.1 inch color display and has a standard configuration of a 3 or 5 lead ECG, Masimo SET SpO2, non-invasive blood pressure, respiration, continuous temperature and IV Drug Calculations. Optional digital displays are provided for invasive blood pressure (up to four) anesthetic agents, O2 and NO2, ST, and CO2.
Status
Terminated
Initiated
2005-09-07
Posted
2005-11-23
Terminated
2007-06-06
Reason
Software anomalies which include shut down while printing, invalid diastolic blood pressure display, and incomplete and inaccurate interpretation reports of 12-Lead ECG data.
Root cause
Other
Action
Datascope sent out a Field Correction notice on 9/7/2005 to all Hospital administrators who have the monitors. This was sent out via certified mail, return receipt requested. A Datascope Service Rep will conduct and document the corrective action during a visit to the customer.
Quantity
2318 monitors
Distribution
The monitors have been distributed domestically and internationally to hospitals and clinics. There are 10 Passport 2 monitors distributed to Naval Hospital, Camp Pendleton, CA. There are 12 Spectrum monitors distributed to two US Govt facilities: VA Medical Center, Hines, IL and VA Medical Center, Syracuse, NY.
Lot, serial or model codes
All Spectrum Monitors that have the following Software Versions: A.19, A.20, B.23, B.24, B.25, B.27, C.18, D.07, E.07, F.09, F.16, F.17, F.26.
510(k) numbers
["K031849"]
PMA numbers
not on file
Product code
DRT
Firm FEI
2221819
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register