Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Magnetic Resonance Imaging. Brand names: Concerto, Concerto Upgrade, Syngo Harmony, Syngo Harmony 087S, Syngo Harmony Upgrade, Syngo Sonata, Syngo Sonata Upgrade, Syngo Symphony, Syngo Symphony Q, and Syngo Symphony Upgrade.

Recall number
Z-0175-03
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Magnetic Resonance Imaging. Brand names: Concerto, Concerto Upgrade, Syngo Harmony, Syngo Harmony 087S, Syngo Harmony Upgrade, Syngo Sonata, Syngo Sonata Upgrade, Syngo Symphony, Syngo Symphony Q, and Syngo Symphony Upgrade.
Status
Terminated
Initiated
2003-10-16
Posted
2003-12-03
Terminated
2004-08-25
Reason
Equipment SAR monitor does not accept new patient data, incorrect RF levels may be administered to the patient.
Root cause
Other
Action
The recalling firm field representatives are visiting each location to correct the problem through a software upgrade.
Quantity
118 units
Distribution
The products were shipped nationwide to hospitals and medical imagining centers.
Lot, serial or model codes
Model numbers 4772906, 7106995, 7104693, 5751438, 7106714, 7104719, 7106425, 7104594, 4772971, and 7106557.
510(k) numbers
["K020991"]
PMA numbers
not on file
Product code
LNH
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register