Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems (Cleveland) Inc: Brilliance 64 Computed Tomography X-ray systems. Intended to produce cross-sectional images of the body.

Recall number
Z-0174-2016
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Brilliance 64 Computed Tomography X-ray systems. Intended to produce cross-sectional images of the body.
Status
Terminated
Initiated
2015-09-14
Posted
not on file
Terminated
2018-06-26
Reason
The firm discovered Ring/Dot artifact due to X-ray measurement error.
Root cause
Software design
Action
On 9-23-2015 the firm sent URGENT Field Safety Notices to their customers. The letter identified the products, problem, and actions to be taken by the customers.
Quantity
108 units
Distribution
Worldwide Distribution-US (nationwide) including DC and the states of AL, AR, AS, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NH, NJ, NV, NY, OH, OK, OR, SC, TN, TX, UT, VA, VT, WI & WV, and the countries of Albania, Algeria, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Italy, Japan, Jordan, Kazakhstan, Latvia, Lebanon, Libya, Lithuania, Malaysia, Mexico, Mongolia, Myanmar, Netherlands, Nigeria, Norway, Oman, Panama, Philippines, Poland, Portugal, Romania, Russia, Rwanda, Samoa, Saudi Arabia, Singapore, South Korea, Spain, State of Palestine, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan & Viet Nam.
Lot, serial or model codes
Model No. 728231; S/N: 2, 9030, 9047, 9050, 9152, 9168, 9234, 9250, 9542, 9552, 9627, 9664, 9690, 9801, 9915, 9996, 9999, 10009, 10066, 10196, 10804, 10807, 29009, 29039, 29047, 29049, 29051, 29070, 29135, 29151, 90035, 90056, 90069, 90085, 90156, 90189, 90192, 95054, 95120, 95138, 95164, 95259, 95285, 95345, 95350, 95359, 95403, 95441, 95442, 95470, 95480, 95497, 95498, 95502, 95517, 95521, 95522, 95561, 95633, 95643, 95656, 95695, 95884, 95885, 95886, 95889, 95893, 95894, 95896, 95897, 95898, 95899, 95900, 95901, 95902, 95903, 95904, 95905, 95906, 95907, 95908, 95909, 95910, 95911, 95912, 95913, 95914, 95915, 95916, 95917, 95918, 95919, 95920, 95921, 95922, 95923, 95924, 95925, 95926, 95927, 95928, 95929, 95930, 95931, 95933, 95934, 95936, 95937, 95938, 95940, 95941, 95942, 95944 & 95946.
510(k) numbers
["K033326"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register