Straumann USA, LLC: Straumann Bone Level Implant 04.1 mm RC, SLActive 8mm, Roxolid, Loxim Article Number: 021.4308 Product Usage: Dental implants are intended for surgical placement in maxillary and/or mandibular arches to provide support for prosthetic restorations in edentulous or partially edentulous patients. Dental implants are for single-stage or two-stage surgery. Dental implants are intended for immediate placement and function on single-tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function. Multiple tooth applications may be rigidly splinted. In the case of edentulous patients, four or more implants must be used
- Recall number
- Z-0174-2015
- Recalling firm
- Straumann USA, LLC
- Product
- Straumann Bone Level Implant 04.1 mm RC, SLActive 8mm, Roxolid, Loxim Article Number: 021.4308 Product Usage: Dental implants are intended for surgical placement in maxillary and/or mandibular arches to provide support for prosthetic restorations in edentulous or partially edentulous patients. Dental implants are for single-stage or two-stage surgery. Dental implants are intended for immediate placement and function on single-tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function. Multiple tooth applications may be rigidly splinted. In the case of edentulous patients, four or more implants must be used
- Status
- Terminated
- Initiated
- 2014-09-24
- Posted
- 2014-11-06
- Terminated
- 2017-06-26
- Reason
- Product transfer piece was fitted with a Narrow Connect (NC) instead of a Regular Connect (RC)
- Root cause
- Packaging process control
- Action
- Straumann USA sent a Field Corrective Action Notice letter dated September 24, 2014 and delivered via United Parcel Service. The letter identified the affected product, problem and actions to be taken. Straumann is requesting the return of unused implants. Returns and Customer Confirmation form is to be returned. Contact firm 781-747- 2514.
- Quantity
- 111 units
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- Lot number: GY256
- 510(k) numbers
- ["K121131"]
- PMA numbers
- not on file
- Product code
- DZE
- Firm FEI
- 1000121052
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28