Skip to the content

FDA Medical Device Recalls (openFDA)

Straumann USA, LLC: Straumann Bone Level Implant 04.1 mm RC, SLActive 8mm, Roxolid, Loxim Article Number: 021.4308 Product Usage: Dental implants are intended for surgical placement in maxillary and/or mandibular arches to provide support for prosthetic restorations in edentulous or partially edentulous patients. Dental implants are for single-stage or two-stage surgery. Dental implants are intended for immediate placement and function on single-tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function. Multiple tooth applications may be rigidly splinted. In the case of edentulous patients, four or more implants must be used

Recall number
Z-0174-2015
Recalling firm
Straumann USA, LLC
Product
Straumann Bone Level Implant 04.1 mm RC, SLActive 8mm, Roxolid, Loxim Article Number: 021.4308 Product Usage: Dental implants are intended for surgical placement in maxillary and/or mandibular arches to provide support for prosthetic restorations in edentulous or partially edentulous patients. Dental implants are for single-stage or two-stage surgery. Dental implants are intended for immediate placement and function on single-tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function. Multiple tooth applications may be rigidly splinted. In the case of edentulous patients, four or more implants must be used
Status
Terminated
Initiated
2014-09-24
Posted
2014-11-06
Terminated
2017-06-26
Reason
Product transfer piece was fitted with a Narrow Connect (NC) instead of a Regular Connect (RC)
Root cause
Packaging process control
Action
Straumann USA sent a Field Corrective Action Notice letter dated September 24, 2014 and delivered via United Parcel Service. The letter identified the affected product, problem and actions to be taken. Straumann is requesting the return of unused implants. Returns and Customer Confirmation form is to be returned. Contact firm 781-747- 2514.
Quantity
111 units
Distribution
US Nationwide Distribution
Lot, serial or model codes
Lot number: GY256
510(k) numbers
["K121131"]
PMA numbers
not on file
Product code
DZE
Firm FEI
1000121052
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register