FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: Syngo Imaging XS VA60 and VA70 in combination with a NAS or HSM license
- Recall number
- Z-0174-2010
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- Syngo Imaging XS VA60 and VA70 in combination with a NAS or HSM license
- Status
- Terminated
- Initiated
- 2009-09-30
- Posted
- 2009-11-10
- Terminated
- 2012-07-05
- Reason
- Incomplete amount of image data; specifically, The image count in the series table of the online and archive database will display the full amount of images received at the Syngo Imaging XS as archived. However, the zip file on the NAS or HSM will contain fewer images-those which have been archived until the time of the error/interruption.
- Root cause
- Software design
- Action
- The recalling firm issued a "Customer Safety Advisory Notice" dated 9/14/09 to all affected customers via Update Instructions IM101/09/S. The notice informs the customer of the potential issue and provides instructions to avoid its occurrence. Questions regarding this issue are directed to the local SIEMENS Uptime Service Center.
- Quantity
- 2 units
- Distribution
- The product was shipped to medical facilities in NY and OH
- Lot, serial or model codes
- Model number 7502029 - serial number 14086 and Model number 10496279 - serial number 1418
- 510(k) numbers
- ["K082430"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28