Skip to the content

FDA Medical Device Recalls (openFDA)

American Contract Systems Inc.: Medical convenience kits Item number/ Product description AHLD15X LABOR DELIVERY PACK - 213387 AKDV96B VAGINAL DELIVERY AKDV96B VAGINAL DELIVERY EVBH28D BIRTHING ROOM PACK UPVD01K VAGINAL DELIVERY PACK

Recall number
Z-0173-2026
Recalling firm
American Contract Systems Inc.
Product
Medical convenience kits Item number/ Product description AHLD15X LABOR DELIVERY PACK - 213387 AKDV96B VAGINAL DELIVERY AKDV96B VAGINAL DELIVERY EVBH28D BIRTHING ROOM PACK UPVD01K VAGINAL DELIVERY PACK
Status
Open, Classified
Initiated
2025-09-04
Posted
2025-10-16
Terminated
not on file
Reason
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Root cause
Nonconforming Material/Component
Action
Firm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers. Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users.
Quantity
32,433 total
Distribution
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Lot, serial or model codes
AHLD15X UDI-DI 191072224767 Lot 8188711 Bag serial no. 17555039 17555044 17555049 17609166 17609181; AKDV96B UDI-DI 191072224019 Lot 7311311 Bag serial no. 47140875 47141919; EVBH28D UDI-DI 191072239051 Lot 8424611 Bag serial no. 17754395; UPVD01K UDI-DI 191072238993 Lot 8439111 Bag serial no. 17682598 Log 8499211 Bag serial no. 17742543
510(k) numbers
not on file
PMA numbers
not on file
Product code
MLS
Firm FEI
3007069947
Firm city
Grimes
Firm state
IA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register