Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Laboratories, Inc: AxSYM Analyzer, Small Heater Block Clamp, manufactured by Abbott Laboratories, Irving, Texas

Recall number
Z-0173-04
Recalling firm
Abbott Laboratories, Inc
Product
AxSYM Analyzer, Small Heater Block Clamp, manufactured by Abbott Laboratories, Irving, Texas
Status
Terminated
Initiated
2002-06-06
Posted
2004-07-20
Terminated
2004-08-09
Reason
Clamp can cause misaligned seating of heater block resulting in inaccurate dispension of MEIA Wash Buffer or MUP into the reaction cell. Causes inconsistent results.
Root cause
Other
Action
All consignees received Technical Service Bulletin (TSB) #83-081. This TSB gave instructions on how to install the AxSYM Small Heater Block clamp upgrade implemented June 6, 2002.
Quantity
15437 units worldwide
Distribution
Nationwide and to the following foreign locations: Mexico, Viet Nam, Venezuela, Columbia, Argentina, Chile, Japan, Austria, Denmark, France, Israel, Norway, Saudi Arabia, Spain, Turkey, India, Phillippines, Brazil, Puerto Rico, Dominicam Republic, Canada, Germany, Taiwan, Thailand, Belgium, Egypt, Greece, Italy, Poland, South Africa, Sweden, United Arab Emirates, Malaysia, Singapore, Hong Kong, Australia, New Zealand, China, Uruguay, Korea, England, Czech Republic, Finland, Hungary, Netherlands, Portugal, Eastern Europe, Switzerland, Central America, and Pakistan
Lot, serial or model codes
All Serial Numbers
510(k) numbers
["K974651"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1628664
Firm city
Irving
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register